Nuclear Magnetic Resonance Instrumentation

Nuclear Magnetic Resonance and Electron Spin Resonance Spectroscopy

  1. Shi Bai

Published Online: 29 SEP 2008

DOI: 10.1002/9780470027318.a6108.pub2

Encyclopedia of Analytical Chemistry

Encyclopedia of Analytical Chemistry

How to Cite

Bai, S. 2008. Nuclear Magnetic Resonance Instrumentation. Encyclopedia of Analytical Chemistry. .

Author Information

  1. University of Delaware, Newark, DE, USA

Publication History

  1. Published Online: 29 SEP 2008

This is not the most recent version of the article. View current version (15 JUN 2016)

Abstract

The nuclear magnetic resonance (NMR) spectrometer consists of a superconducting magnet, a sample probe, a radio-frequency (RF) transmitter, a receiver, and a computer for instrument control and data processing. The superconducting magnet provides an ultrastable magnetic field in the central volume of the magnet. To maintain the current, the solenoid that produces the field is immersed in liquid helium to maintain the superconducting state. Surrounding the magnet volume is a set of vacuum jackets and a liquid nitrogen–filled volume to isolate the low-temperature regions from room temperature. The magnet has a bore to house the sample probe and a room-temperature shim (RTS) coil assembly to reduce inhomogeneity of the magnetic field across the active sample volume. The sample to be analyzed is usually dissolved in a deuterated solvent and is contained in a specially ground glass tube that is inserted into the volume of the probe coil. RF pulses, from one or a set of high-power transmitters, are applied to the sample through the RF coil that surrounds the active sample area to perturb the sample. The RF pulses are controlled to ensure precise frequency, phase, duration, and amplitude of the RF voltage. The response of the sample is gathered by recording the induced voltage across the RF coil. The response (the signal) is digitized as a so-called free induction decay (FID) with a high-performance RF receiver including a preamplifier, an amplifier, and an analog-to-digital converter (ADC). By using Fourier transformation (FT), the time-domain FID is converted into the frequency-domain spectrum, often known as the NMR spectrum.