Falling coaxial cylinder viscometry of polymer melts

Authors


Abstract

This article considers possible erxors in experiment or analysis and alternative methods of data treatment for a falling coaxial cylinder viscometer. The viscometer uses a weighted rod falling into a polymer melt contained in a closed end concentric cylinder. Flow relations can be solved for non-Newtonian fluid response in steady, simple shearing flow. Although the rod fall is found not to be steady, as assumed, this produces negligible error in the computed flow curves. Consistent flow curves which were independent of apparatus geometry were obtained by a method in which instantaneous rod velocities were estimated graphically from raw data and an end correction was taken into account.

Abstract

On considére, dans le présent travail, les erreurs qui peuvent se glisser dans les experiences ou analyses et la mèthodes alternatives du traitement des données dans le cas d'un viscosimétre basé sur la chute d'un cylindre coaxial. Dam le dit viscosimétre, on laisse tomber une tige alourdie par un poids dans un polymére en fusion qui se trouve dans un cylindre fermé et concentrique. On peut résoudre les équations d'écoulement relatives à la réaction d'un fluide non-newtonien engagé dans un écoulement stable et à cisaillement simple. Bien qu'on constate que la chute de la tige n'est pas aussi régulière qu'on le supposait, l'erreur résultante dans les courbes d'écoulement calculées est négligeable. On a obtenu des courbes consistantes d'écoulement, qui ne dépendaient pas de la géométrie de l'appareillage, en utilisant une méthode dans laquelle on a évalué graphiquement les vélocites instantanées de la tige, à partir de données brutes, et tenu compte d'une correction finale.