Bubble Dynamics Measurements Using Four-Point Optical Probe

Authors


Abstract

By conducting an analysis of the measurement errors it was found that bubble dynamics obtained by two-point probes can be highly inaccurate. A four-point optical probe with appropriate data processing algorithm has been developed so that the bubble velocity vector can be obtained. This probe provides directly the value of specific gas-liquid interfacial area. However, for obtaining accurate values of bubble velocity and size, the four-point probe must be calibrated by a CCD camera at the same conditions. Calibration columns have been developed for this purpose where the bubble size and velocity can be varied to calibrate the measurements by the four-point optical probe against a CCD camera.

Abstract

En effectuant une analyse des erreurs de mesure, on a trouvé que la dynamique des bulles obtenue au moyen de sondes à deux points peut être très imprécise. On a donc mis au point une sonde optique à quatre points avec un algorithme de traitement des données approprié de façon à obtenir le vecteur de vitesse des bulles. Cette sonde donne directement la valeur de la région interfaciale gaz-liquide spécifique. Cependant, pour obtenir des valeurs précises de la vitesse et de la taille des bulles, la sonde à quatre points doit être calibrée au moyen d'une caméra CCD avec les mêmes conditions. Des colonnes de calibration ont été mises au point à cette fin là, qui permettent de faire varier la taille et la vitesse des bulles pour calibrer les mesures de la sonde optique à quatre points par rapport à la caméra CCD.