A Fully Continuous Supercritical Fluid Extraction System for Contaminated Soil

Authors


Abstract

A fully continuous lab scale supercritical fluid extraction system has been designed, built and tested. The system continuously pumps carbon dioxide under supercritical conditions and soil slurry into a counter-current contacting column. Delhi Loamy Sand, spiked with approximately 10 mg/g of naphthalene, was used as the testing soil. The soil slurry ranged from 0.0028 g dry soil per g slurry to 0.072 g/g. The operating temperature was 43°C and the operating pressure was 7.7 MPa. Near steady state, fully continuous flow was achieved with runs lasting as long as 2 h. All carbon dioxide samples contained measurable quantities of naphthalene and the measured mass transfer coefficients were as high as 4.6×10−4 s−1.

Abstract

On a conçu, construit et testé un système d'extraction par fluide supercritique à l'échelle de laboratoire entièrement continu. Le système pompe en continu du dioxyde de carbone dans des conditions supercritiques et des suspensions de sol dans une colonne de contact à contre-courant. Du sable limoneux de Dehli, reconstitué avec environ 10 mg/g de naphtalène, a été utilisé comme sol d'essai. La suspension de sol varie de 0,0028 g de sol sec par g de suspensions à 0,072 g/g. La température de fonctionnement est de 43ºC et la pression de fonctionnement de 7,7 MPa. Un écoulement entièrement continu à l'état presque permanent a été obtenu avec des essais d'une durée pouvant atteindre 2 heures. Tous les échantillons de dioxyde de carbone contiennent des quantités mesurables de naphtalène et les coefficients de transfert de matière mesurés peuvent atteindre 4,6×10−4s−1.