Two accurate methods to obtain the spectral sensitivity of a Raman spectrometer device

Authors

  • R. Ouillon,

    1. Département de Recherches Physiques, Laboratoire associé au CNRS n°71, Université P. et M. Curie, 75230 Paris Cedex 05, France
    Search for more papers by this author
    • Research supported in part by the Community Bureau of Reference of the European Communities under operating grant 494/1/4/41/79/8 BCR-F.

  • S. Adam

    1. Département de Recherches Physiques, Laboratoire associé au CNRS n°71, Université P. et M. Curie, 75230 Paris Cedex 05, France
    Search for more papers by this author

Abstract

We have compared the spectral transmission coefficient of a Raman spectrometer and photomultiplier device determined (a) from a continuous spectrum (calibrated tungsten lamp) and (b) from the spectrum of the rotational Raman lines of D2 molecules. The two methods give a similar sensitivity curve as a function of wavelength. However the first method is more accurate in the determination of the spectral response over a wide spectral range (˜3000 cm−1 or more); and the second method is better than the first for obtaining accurate value of the intensity of the Raman lines and their absolute scattering cross sections.