Development of a compact MRI system for trabecular bone volume fraction measurements

Authors


Abstract

A compact MRI system for measuring trabecular bone volume fraction (TBVF) of the calcaneus was developed with the use of a 0.21 T permanent magnet and portable MRI console. The entire system weighed < 600 kg and was installed in a 2 m × 2 m space. Two cross-sectional spin-echo images of a heel acquired with external reference phantoms (total measurement time = 5 min) were used to quantify the TBVF of the calcaneus. The linearity and reproducibility of the measurements were evaluated by means of proton density-adjusted phantoms. Comparative measurements with quantitative ultrasound (QUS) in groups of healthy female volunteers showed a relatively high positive correlation (R2 = 0.4539, 0.2693) between TBVF and the speed of sound (SOS). These results demonstrate the potential of this new system for measuring bone density. Magn Reson Med 52:440–444, 2004. © 2004 Wiley-Liss, Inc.