Keywords:

  • diffusion imaging;
  • motion;
  • signal loss partial k-space sampling;
  • homodyne reconstruction

Abstract

Partial k-space sampling is frequently used in single-shot diffusion-weighted echo-planar imaging (DW-EPI) to reduce the TE and thereby improve the SNR. However, it increases the sensitivity of the technique to bulk rotational motion, which introduces a phase gradient across the tissue that shifts the echo in k-space. If the echo is displaced into the high spatial frequencies, conventional homodyne reconstruction fails, causing intensity oscillations across the image. Zero-padding, on the other hand, compromises the image resolution and may cause truncation artifacts. We present an adaptive version of the homodyne algorithm that detects the location of the echo in k-space and adjusts the center and width of the homodyne filters accordingly. The adaptive algorithm produces artifact-free images when the echo is shifted into the high positive k-space range, and reduces to the standard homodyne algorithm in the absence of bulk motion. Magn Reson Med 57:614–619, 2007. © 2007 Wiley-Liss, Inc.