Correction for artifacts induced by B0 and B1 field inhomogeneities in pH-sensitive chemical exchange saturation transfer (CEST) imaging

Authors

  • Phillip Zhe Sun,

    Corresponding author
    1. Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Charlestown, Massachusetts, USA
    • Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Charlestown, MA 02129
    Search for more papers by this author
  • Christian T. Farrar,

    1. Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Charlestown, Massachusetts, USA
    Search for more papers by this author
  • A. Gregory Sorensen

    1. Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Harvard Medical School, Charlestown, Massachusetts, USA
    Search for more papers by this author

Abstract

Chemical exchange saturation transfer (CEST) imaging provides an indirect detection mechanism that allows quantification of certain labile groups unobservable using conventional MRI. Recently, amide proton transfer (APT) imaging, a variant form of CEST imaging, has been shown capable of detecting lactic acidosis during acute ischemia, providing information complementary to that of perfusion and diffusion MRI. However, CEST contrast is usually small, and therefore, it is important to optimize experimental conditions for reliable and quantitative CEST imaging. In particular, CEST imaging is sensitive to B0 and B1 field, while on the other hand; field inhomogeneities persist despite recent advances in magnet technologies, especially for in vivo imaging at high fields. Consequently, correction algorithms that can compensate for field inhomogeneity-induced measurement errors in CEST imaging might be very useful. In this study, the dependence of CEST contrast on field distribution was solved and a correction algorithm was developed to compensate for field inhomogeneity-induced CEST imaging artifacts. In addition, the proposed algorithm was verified with both numerical simulation and experimental measurements, and showed nearly complete correction of CEST imaging measurement errors caused by moderate field inhomogeneity. Magn Reson Med, 2007. © 2007 Wiley-Liss, Inc.