Independent estimation of T*2 for water and fat for improved accuracy of fat quantification

Authors


Abstract

Noninvasive biomarkers of intracellular accumulation of fat within the liver (hepatic steatosis) are urgently needed for detection and quantitative grading of nonalcoholic fatty liver disease, the most common cause of chronic liver disease in the United States. Accurate quantification of fat with MRI is challenging due the presence of several confounding factors, including T*2 decay. The specific purpose of this work is to quantify the impact of T*2 decay and develop a multiexponential T*2 correction method for improved accuracy of fat quantification, relaxing assumptions made by previous T*2 correction methods. A modified Gauss-Newton algorithm is used to estimate the T*2 for water and fat independently. Improved quantification of fat is demonstrated, with independent estimation of T*2 for water and fat using phantom experiments. The tradeoffs in algorithm stability and accuracy between multiexponential and single exponential techniques are discussed. Magn Reson Med 63:849–857, 2010. © 2010 Wiley-Liss, Inc.

Nonalcoholic fatty liver disease (NAFLD) is now recognized as the most common cause of chronic liver disease, afflicting up to 30% of all Americans (1). It is an emerging condition closely related to obesity and insulin resistance. Importantly, NAFLD's prevalence among children is reported to be up to 10% overall, and as high as 53% in obese children (2–4). NAFLD is expected by many experts to become a leading cause of end-stage liver disease as the prevalence of obesity increases in the general population, both in the US and worldwide.

The hallmark feature of NAFLD is intracellular accumulation of triglycerides within hepatocytes (steatosis). In many patients, steatosis leads to inflammation and fibrosis, and ultimately to cirrhosis, with subsequent liver failure or development of hepatocellular carcinoma. In such patients, liver transplant is the only definitive option for cure. Nontargeted liver biopsy, which is the current gold standard for diagnosis of NAFLD, is limited by its high cost, morbidity, and importantly, its high sampling variability due to the heterogeneous nature of intracellular lipid accumulation. Quantitative assessment of liver fat using MRI is attractive because it can assess fat over the entire liver, thus avoiding the sampling variability, as well as the risks and high cost of biopsy.

Chemical shift–based MRI techniques are currently under development by many groups for the quantification of liver fat (5–10). These methods exploit the differences in chemical shift between water and fat (3.29 ppm between water and the main resonance peak of fat, 210 Hz at 1.5 T). Chemical shift–based methods are often used to estimate the concentration of triglycerides through the use of the fat fraction, which is independent of amplitude of radiofrequency field coil sensitivities and therefore is a useful metric of fat concentration (11).

Two-point methods acquire two images, one in which water and main peak of fat are in phase and the other in which they are out of phase (12, 13). Multipoint chemical shift–based methods (5,9,14,15) separate the signals of water and fat, even in the presence of magnetic field inhomogeneities, permitting estimation of fat fractions over a full dynamic range of 0–100%. However, two-point and multipoint chemical shift–based water-fat separation methods are limited for fat quantification because of T*2 decay, T1-related bias (7, 16), and insufficient spectral determination of fat, which has multiple spectral peaks (7, 17), which leads to inaccurate separation of water and fat signals (17). In addition, the recombination of magnitude fat and water images into a fat-fraction image can also introduce noise-related bias (16).

These confounding factors have recently been addressed by several groups, including small flip angle and dual flip angle approaches to avoid T1-related bias (7, 16), magnitude discrimination and phase constrained methods to avoid noise bias (16), and the use of accurate spectral modeling to separate fat signal more accurately (7, 17).

T*2 decay is well known to corrupt estimates of fat fraction and is particularly important in chronic liver diseases such as NAFLD, where concomitant iron overload can occur in up to 40% of patients (18, 19). Typical values of T*2 in the livers of healthy individuals are variable but typically exceed 20 ms (20). In the case of patients with iron overload, however, significant T*2 shortening may exist. In such cases, T*2 can be less than a few milliseconds (21). If T*2 decay is not included in the signal model, it will corrupt the accuracy of all chemical shift–based fat quantification methods. Yu et al. (22) and Bydder et al. (7) independently introduced methods that estimate T*2 from the signal and demodulate its effects, correcting estimates of fat fraction.

Yu et al. (22) assume T*2 decay for water and fat signals to be identical, and Bydder et al. (7) estimate only a single T*2 value independently. Even though Bydder et al. (7) allow T*2 values of water and fat to differ, the estimated value of T*2 for fat is dependent on the estimated T*2 value of water by the constraint that restricts the T*2 of fat to be always smaller than the T*2 of water. Therefore, the estimated values of T*2 of water and fat are not independent. While correcting for a single T*2 value has been shown to improve estimates of fat fraction (7,10,22), the assumption that T*2 of water and fat is interdependent is questionable and could lead to inaccuracies in the estimation of fat fraction. We explore the inaccuracies that can result from this simplification and propose a new correction algorithm with independent T*2 modeling for water and fat, with the goal of improving the accuracy of fat-fraction estimation. The accuracy and stability of this algorithm are compared with that of a single T*2 correction method.

THEORY

Signal Model

The signal from a voxel containing water and fat with independent T*2 decay for all fat peaks can be written as:

equation image(1)

Here, W and F are the water and fat signals, ψ is the shift (hertz) in the spectrum caused by local amplitude of static field inhomogeneities. R*2,w is the R*2 of water. Δfp and R*2,fp are the central resonance frequency and R*2 of the pth fat peak, respectively. rp is the relative proportion of the pth fat peak such that equation imagerp = 1. Note that both the frequencies (Δfp) and relative amplitudes (rp) of the fat peaks are assumed to be known (5, 17).

Equation 1 is a multiple T*2 signal equation, where each fat peak has a different T*2 value. If we simplify this expression by considering T*2 of water to be the same as the T*2 of all fat peaks, i.e., R*2,w = R*math image = Rmath image, then we obtain the single T*2 signal model used by Yu et al. (17, 22):

equation image(2)

In the Yu et al. (17) single T*2 method, the fat spectrum was known a priori, either measured using MR spectroscopy or estimated directly from the data using spectrum self-calibration algorithms (17).

Fat Quantification Without Dual T2* Correction

The correction of a single T*2 is known to improve estimates of fat fraction (7, 17). Particularly, when fat fractions to be estimated are low, and when T*2 of water and fat are similar, single T*2 methods can avoid large errors. To examine the importance of T*2 correction, simulations were conducted to find absolute percentage errors in the estimation of fat fraction with signal models that do not account for T*2 decay.

Simulations that examined the apparent fat fraction obtained using two-point in-phase/out-of-phase (IOP) imaging (12, 13) without T*2 correction are shown in Fig. 1a. As is seen in this plot, errors as large as 30% can occur. In addition, three-point methods such as three-point IDEAL (Iterative Decomposition of water and fat with Echo Asymmetry and Least squares estimation) (9, 23) can generate errors up to 15% if no T*2 correction is performed (Fig. 1b). In this simulation, we have assumed that the T*2 of water and fat are equal. In Fig. 1, each different line represents the estimated fat fraction from either two-point (Fig. 1a) or three-point (Fig. 1b) fat-estimation methods that do not incorporate any T*2 correction, when T*2 of water = T*2 of fat = 25 ms or 15 ms or 10 ms or 5 ms.

Figure 1.

Simulations of the apparent fat fractions estimated using two-point IOP and conventional three-point IDEAL, neither of which corrects for T*2 decay. a: IOP demonstrates large errors with negative, paradoxical values of apparent fat fraction. An apparent fat fraction of −5% occurs even when no fat is present and T*2 is normal (25 ms). b: Relatively large errors are also seen with three-point IDEAL when no correction for T*2 decay is made. These errors are clinically very relevant and underscore the need for T*2 correction. For these simulations, it is assumed that the T*2 of water and fat are equal. These simulations underscore the point that when T*2 of water and fat are equal, single T*2 methods can avoid large errors, particularly when fat fractions are low. Each different line represents the estimated fat fraction from either two-point (a) or three-point (b) fat estimation methods that do not incorporate any T*2 correction, when T*2 of water = T*2 of fat = 25 ms or 15 ms or 10 ms or 5 ms.

Although the correction of a single value of T*2 improves estimates of fat fraction (7, 17), the validity of the assumption that the rates of signal decay of water and fat are the same is unclear as there is no physiologic basis for this assumption. To examine the importance of dual T*2 correction, simulations that demonstrate the impact of different fat and water T*2 values were conducted to find absolute percentage errors in the estimation of fat fraction with signal models that model the T*2 of water and fat to be equal and are shown in Fig. 2.

Figure 2.

Simulations of the percentage error in measured fat fraction when using single T*2 correction with six-point IDEAL to reconstruct data where the T*2 of fat and water is different. Different fat-fractions were simulated: (a) 10%, (b) 20%, (c) 30%, and (d) 50%. Zero percent fat fraction is not shown because single T*2 methods generate no error for this case. The labels on the contour plots show the absolute percentage errors in fat-fraction estimation. The error in the apparent fat fraction is zero along the diagonal of the contour plots because the single T*2 correction method accurately removes the error caused by T*2 decay. Relatively large errors, however, can occur when T*2 of water and fat are not equal, particularly at higher fat-fraction and shorter T*2 values (longer R*2).

When the values of T*2 for water and fat are identical (along the diagonal), this model accurately measures fat fraction and there is no error. However, when T*2 of water and fat are not the same, errors in the estimation of fat fraction using a single T*2 correction method can exceed 20% when T*2 values of water and fat are short (Fig. 2), particularly at higher fat fractions.

In the simulations for both Figs. 1 and 2, it was assumed that ψ has been accurately demodulated from the signals. For given values of W, F, and corresponding T*2's, Eqs. 2 and 3 were used to generate simulated signals, and the two-point IOP, three-point IDEAL, and single T*2 methods were used to estimate the apparent fat fraction from the simulated data. The difference between the estimated fat fractions and the true fat fractions is reported in Figs. 1 and 2.

Dual T2* Method for Independent Estimation of T2* of Water and Fat

Clearly there is a need for correction of T*2 decay with MRI methods attempting to quantify fat. Further, correction for T*2 decay that assumes a common rate of signal decay for water and fat can lead to very large errors in the estimated fat fraction if the rates of decay are in fact significantly different and there is also a large amount of fat. For this reason, we develop a signal model and estimation algorithm below, which permits independent estimation of T*2 decay rates for water and fat to improve the accuracy of fat quantification with MRI.

Because all the protons on a single triglyceride molecule experience very similar microscopic magnetic field inhomogeneities, it may be reasonable to assume that all the fat peaks have the same T*2 values, although these values are different from that of water. We define the variable R*2,f such that R*2,f = R*math image for all p. Equation 1 can now be written

equation image(3)

We will use Eq. 3 as the signal model for the dual T*2 estimation described in the remainder of this work. If we define Rw = R*2,w and Rf = R*2,f , then real and imaginary parts of s(t), namely, sr(t) and si(t), can be written

equation image(4)
equation image(5)

From Eqs. 4 and 5, the seven parameters that must be estimated are Wr, Wi, Fr, Fi, R*2,w, R*2,f and ψ, where Wr, Wi and Fr, Fi are the real and imaginary parts of W and F, respectively. A minimum of four complex images (equivalent of eight measurements) must be acquired at different echo times to estimate these parameters. We propose an iterative technique based on the Gauss-Newton method for multiple variables to independently estimate these parameters.

The initial iteration of the algorithm uses starting values estimated from a single exponential T*2 correction method such as T*2-IDEAL (17, 22). Initial guesses for Wr, Wi, Fr, Fi, R*2,w = R*2,f = R*2, and ψ from a single T*2 correction method are useful starting values not only to reduce the number of iterations but also to avoid convergence to local minima (24).

Estimates of the parameters are subsequently updated using Taylor's first-order approximation for multiple variables, as is done in the Gauss-Newton method. The difference between the measured signal and the signal calculated from the parameters in the current iteration (Eqs. 4 and 5) is then calculated. This difference was reduced by finding a constant multiplying factor that minimizes the L2 norm for the difference vector used to update the parameters (25, 26). In this way, the step size obtained from the Gauss-Newton method was optimized by performing a linear search in the direction of the difference vector. The process was repeated until the mean squared value of the step size was reduced to a value smaller than a predetermined value, or if the numbers of iterations exceeded a particular count. The algorithm is described as follows and is summarized in Fig. 3.

Figure 3.

Flow chart that summarizes the algorithm to estimate water and fat, using the signal model that allows independent estimation of T*2 for water and fat.

Notation

Let X = [WrWiFrFiR*2,wR*2,f ψ] be the vector representation of the parameters to be estimated, such that each element of the vector is real. Let equation image be the estimate of the vector X at the jth iteration of the algorithm. If N echoes images are acquired at echo times of t1, t2tntN, then let S be the vector representation of real and imaginary echo points such that

equation image

where sr(tn) and si(tn) are the real and imaginary parts of s(tn), the signal at the nth echo. Let the estimated signal vector calculated using equation image from the signal model in Eqs. 4 and 5 be equation image, where ${\hat {\bf S}}_{\rm j} = \left[ {\matrix{ {s_j^r \left( {t_1 } \right)} & {s_j^r \left( {t_2 } \right)} & {...} & {\matrix{ {s_j^r \left( {t_N } \right)} & {s_j^i \left( {t_1 } \right)} & {s_j^i \left( {t_2 } \right)} & \quad {...}\quad & {s_j^i \left( {t_N } \right)} \cr } } \cr } } \right]^T$.

Let equation image be the correction factor for equation image to reduce the mean squared error between measured signal vector S and the estimated signal vector equation image. The algorithm becomes:

1. Initialization

The initial guess for all the parameters is obtained from a single exponential T*2 correction techniques such as multipeak T*2-IDEAL method (17).

2. Applying Gauss-Newton Method

Next, we iteratively update the estimate of X using Taylor's first-order approximation for multiple variables, i.e.,

equation image(6)

and

equation image(7)

The two expressions in Eqs. 6 and 7 evaluated at different echo times t1, t2, … , tN can be written in the matrix form as

equation image(8)

Here, B is a 2N × 7 matrix whose elements contain the partial derivatives from Eqs. 4 and 5. The expression for the matrix B and explicit expressions for partial derivatives are in the Appendix.

3. Estimating the Step Size

The correction vector for the estimated parameter vector is estimated using the Moore-Penrose pseudoinverse matrix, i.e.,

equation image(9)

4. Linear Search to Optimize Step Size

The step size obtained is optimized (25, 26) by multiplying equation image with a constant multiplying factor k that minimizes the residual error. k is determined by minimizing the L2 norm of equation image. This optimization reduces the number of iterations required before the algorithm converges to a solution (27). When this optimization is not used, the value of k is set to 1.

5. Updating the Estimation

After determining the value of k, we update the estimation of vector equation image for the next iteration, i.e.,

equation image(10)

6. Iteratively Converging to Solution

We repeat the procedure until the step size equation image converges to a very small value or the number of iterations become more than a predefined number. Empirically, we have found that a stopping point of 0.1% of the magnitude of the X, or when the number of iterations exceeds 50, is a useful stopping criterion. Typically, the algorithm converged in 20 iterations.

Estimating T2* of Fat in the Case of Extreme Fat Fractions

If the estimated fat fraction from single exponential techniques is 0% (or 100%), then we have no need for dual exponential technique as there is no fat (water) to estimate. The dual T*2 technique becomes ill conditioned at very low (high) fat fractions when there is very little signal from fat (water) because the very low fat signal observed at the sample times could occur either from the absence of fat or from extremely short T*2 values of fat. In these cases, the B matrix becomes singular, and there are multiple possible solutions to the problem. This problem can be solved by imposing constraints on the estimated values of R*2 of water and fat. We have constrained the values of R*2 of water or fat to be less than 300 sec−1 and greater than 0 sec−1, which encompasses a very wide range that will handle all physiologically possible values of R*2 in water or fat.

MATERIALS AND METHODS

Phantom Construction and Imaging

A fat/water/iron oxide phantom was constructed containing varying known fat fractions (0.0, 0.03, 0.05, 0.11, 0.21, 0.32, 0.42, 0.52) and superparamagnetic iron oxide (SPIO) concentrations (0, 10, 21, 32 μg Fe/mL), with details described elsewhere (28).

Imaging was performed using the head coil of a 1.5-T Signa HDx system (v14.0; TwinSpeed; GE Healthcare, Waukesha, WI) using an investigational multiecho three-dimensional spoiled gradient echo pulse sequence. Imaging parameters included the following: TEmin = 1.4 ms, ΔTE = 1.6 ms, six echoes/TR, and TR = 42.7 ms, with flip = 5° to minimize T1 bias (16), field of view = 35 × 35 cm, matrix = 256 × 256, BW (Bandwidth) = ±100 kHz, one signal average, and slice thickness = 8 mm.

RESULTS

Figure 4 shows the phantom fat-fraction images reconstructed using methods considering no T*2 correction (Fig. 4a), single T*2 correction (Fig. 4b), and dual T*2 correction (Fig. 4c). Improved and uniform estimation of fat fraction can be observed using the proposed method, especially at higher fat-fraction.

Figure 4.

Fat-fraction images from the fat/water/SPIO phantom reconstructed with (a) no T*2 correction, (b) single T*2 correction, and (c) dual T*2 reconstruction. For each row, fat fraction should remain constant as the SPIO concentration increases. Only with the dual T*2 correction method does this occur.

Figure 5 plots the estimated values of fat-fraction at different true fat-fraction (0.0, 0.03, 0.05, 0.11, 0.21, 0.32, 0.42, 0.52) and iron concentrations (0, 10, 21, 32 μg Fe/mL) without T*2 correction (Fig. 4a), single T*2 correction (Fig. 4b), and with T*2 correction using the proposed dual T*2 method (Fig. 4c). Error bars show the standard error of the mean. For iron concentrations of 32 μg/mL, errors in estimated fat fractions were reduced from 30% with no T*2 correction to 25% with single T*2 correction and to less than 5% using dual T*2 correction. As expected, the impact of T*2 correction was less at lower fat fractions and/or lower iron concentrations.

Figure 5.

Fat fraction measured from the fat/water/SPIO phantom reconstructed with (a) no T*2 correction, (b) single T*2 correction, and (c) dual T*2 reconstruction. Error bars show the standard error of the mean. As the SPIO concentration increases, the fat fraction should remain constant if the correction algorithm is removing the effects of T*2 decay correctly. Large errors are seen without T*2 correction, and although the single T*2 correction method improves estimates of fat fraction, relatively large errors are still seen at high fat fractions. Only with the dual T*2 correction method does the estimated fat fraction agree closely with the known fat fraction, independent of SPIO concentration.

Figure 6 shows the estimated R*2 values from single T*2 correction (Fig. 6a) and the dual T*2 correction methods (Fig. 6b,c) at different iron concentrations for different fat fractions. Error bars show the standard error of the mean. These plots demonstrate large differences in the estimated R*2 values of water and fat in this phantom. This is particularly true at higher SPIO concentrations where R*2 of water more rapidly increases, compared to that of fat. The results for R*2 show that the R*2 of water is similar to the R*2 of fat for SPIO concentrations below 5 μg/mL and that the R*2 of water is significantly greater than the R*2 of fat for SPIO concentrations above 5 μg/mL, underscoring the need for a dual T*2 correction. It can also be observed that the estimated R*2 values using the single exponential correction method are much closer to the R*2 values of water from the dual exponential correction technique for low fat fractions and gradually approach the R*2 values of fat using the dual exponential correction method as the fat fraction increases from 11% to 52%. These results also suggest that R*2 values of water and fat are not equal and need to be estimated independently.

Figure 6.

Estimated R*2 values from single T*2 correction method (a) and the dual T*2 correction method (b,c) at increasing SPIO concentrations and different fat fractions. Error bars show the standard error of the mean. With the single T*2 correction method, the estimated values of R*2 increase as SPIO concentrations increase, and there is a strong dependence of the estimated R*2 with fat fraction. Using the dual T*2 correction method, the R*2 of water is more strongly affected by increasing concentrations of SPIO. Interestingly, there is relatively minimal dependence of the estimated R*2 values on fat fraction using the dual T*2 method. These findings suggest that the dual T*2 method more accurately reflects the signal behavior of the fat/water/SPIO phantom.

DISCUSSION

Accurate quantification of fat requires correction for the differing T*2 decay of water and fat. Simulations demonstrate that very large errors in the estimation of fat fraction can occur using IOP imaging or three-point methods that do not correct for T*2 decay. IOP imaging produces negative apparent fat fractions even at normal values of T*2 (fat fraction = −5%) and much larger errors with higher degrees of iron overload (fat fraction = −30%). This phenomenon occurs because the in-phase image is acquired at a longer echo time and therefore has more T*2 weighting. In the absence of fat, a paradoxical value of fat fraction that is less than zero will be calculated. Although correction for a single, shared value of T*2 for water and fat improves errors in the apparent fat fraction, large errors can occur when the T*2 values of water and fat are different, particularly at high fat fractions and short T*2 values.

Errors such as these are clinically relevant. Although the precise concentration of triglyceride that is considered abnormal is debated, it is generally thought that fat fractions above approximately 5% are clinically important. In a large MR spectroscopy study performed by Szczepaniak et al. (29) in 2349 participants of the Dallas Heart Study, they defined a 95th percentile cutoff of 5.56% hepatic fat fraction as abnormal, based on a subset of 345 patients with no identifiable risk factors for hepatic steatosis. Using this cutoff to distinguish normal from abnormal patients, clearly correction for T*2 is essential for any MRI method attempting to quantify fat. This is particularly true for any method used for early detection of steatosis when fat is in low concentration.

In our phantom experiments, relatively large errors occur in the estimation of fat fraction when no T*2 correction is used. The single T*2 correction method reduces the error, but relatively large errors still occur at higher SPIO concentrations when high levels of fat are also present. This occurs because the SPIOs shorten the T*2 of water more than fat at higher iron concentrations, and the assumption that the signals from water and fat have similar decay rates breaks down. Only with an algorithm that allows for independent estimation and correction for T*2 of water and fat can accurate estimates of fat fraction be made in these situations.

The behavior of the estimated R*2 from the phantom experiments explicitly demonstrates the differential effect of SPIO on the water and fat signals. This is clearly seen in the estimated R*2 values of water and fat using the dual T*2 method. This dependence is also seen indirectly with the single T*2 method, where there is a strong dependence on the apparent R*2 value, due to an averaging effect from water and fat, i.e., one would expect that the estimated R*2 using the single T*2 method would be similar to that of water at low fat fractions and closer to the R*2 of fat at higher fat fractions. This behavior was seen and indicates that modeling of independent decay rates for water and fat will improve the accuracy of the signal model.

The differences in R*2 between water and fat estimated using the dual T*2 method indicate that SPIOs preferentially accelerate the signal decay of water more than that of fat. This effect may be caused because SPIOs are soluble in water and insoluble in fat and therefore are more isolated from fat molecules than water. It is unknown whether this phantom accurately reflects the underlying microscopic relationship of water, fat and iron in an iron-overloaded, steatotic liver. Further work is needed to understand whether important differences in R*2 between water and fat occur in vivo, similar to those seen in our phantom experiments.

There are important disadvantages of including independent correction of T*2 of water and fat. The introduction of this additional degree of freedom dramatically increased the complexity of the estimation algorithm, including the need for constrained reconstruction methods to avoid instabilities at low (and high) fat fractions. The computation of the matrix B and its pseudoinverse is computationally very expensive because B has 2N × 7 elements (e.g., 12 × 7 = 84 for six echoes), and for each iteration we must calculate all of the partial derivatives. The current implementation of the algorithm in MATLAB 7.5.0 (R2007b, The Mathworks, Natick, MA) on Dell Latitude D630 (Dell Inc., Round Rock, Texas) laptop with Intel (Santa Clara, CA) Dual core processors, takes about 45 min for generating a single 256 × 256 fat-fraction image with dual T*2 correction. However, this implementation was intended for feasibility and was not optimized for computation time. Previous unoptimized implementations for single T*2-IDEAL took about 30 min for generating a single fat-fraction image. The optimized single T*2-IDEAL method now generates a single fat-fraction image in a few seconds using the on-line implementation. Once the present dual T*2 algorithm is also optimized, we expect it to compute an entire fat-fraction image, similar to that shown in Fig. 4, in a time duration comparable, and at most no more than twice that of what is taken by the optimized single T*2 IDEAL algorithm. In addition, the estimation of additional degrees of freedom is expected to degrade the noise performance of this method. Full evaluation of the noise performance of this method is involved and beyond the scope of the current work. Future work will focus on a full evaluation of the noise performance algorithm optimization.

In conclusion, noninvasive quantification of fat is needed for early detection and grading of fatty liver disease. T*2 values of water and fat are independently estimated with our method using a modified Gauss-Newton method for multiple variables, relaxing the assumptions made by single exponential T*2 correction methods, that assume a common value of T*2 for both water and fat. Improved quantification of fat can be achieved using independent estimation of T*2 values for water and fat. Future work will optimize the noise performance and investigate the performance and importance of this method for in vivo applications.

Acknowledgements

We wish to thank the University of Wisconsin ICTR, funded through an NIH Clinical and Translational Science Award, grant number 1UL1RR025011, NIH Grant RC1EB010384, and GE Healthcare for their support.

APPENDIX

The matrix B is a 2N × 7 matrix determined by the partial derivatives shown in Eqs. 6 and 7 calculated at different time points, tn:

equation image(A.1)

Elements of this matrix must be calculated for each iteration. The partial derivatives contained in the elements of B are listed below:

equation image(A.2)
equation image(A.3)
equation image(A.4)
equation image(A.5)
equation image(A.6)
equation image(A.7)
equation image(A.8)
equation image(A.9)
equation image(A.10)
equation image(A.11)
equation image(A.12)
equation image(A.13)
equation image(A.14)
equation image(A.15)