Objectives

The main goal was to develop a reproducible method for estimating the diffusion of water in human fetal lung tissue using diffusion-weighted imaging (DWI). A secondary objective was to determine the relationship of the apparent diffusion coefficients (ADCs) in the fetal lung to menstrual age and total lung volume.

Methods

Normal pregnant volunteers were scanned on a 1.5-Tesla (T) magnetic resonance imaging (MRI) system. The MRI system was equipped with 40-mT/m gradients (slew rate 200 T/m/s, rise time 0.2 ms). A six-channel body array coil was used for signal reception. Single-shot DWI utilized TE/TR 125/3400 ms, slice thickness 4 mm, field of view 280 mm × 280 mm, interslice gap 0.8 mm and a matrix of 128 × 128. The voxel size was 2.5 mm × 2.5 mm × 4.0 mm. Two b-values (0 and 1000) were chosen along three orthogonal directions. ADC maps were created using assigned b-values. Simple linear regression was performed with Pearson correlation coefficient. Interexaminer and intraexaminer bias, and 95% limits of agreement (LOA) were determined using Bland–Altman plots.

Results

Forty-seven scans were performed at a mean ± SD of 29.2 ± 4.5 weeks. The median coefficient of variation for ADC was 5.6% (interquartile range, 4.0–8.1%). No differences in ADC values were found between right and left lungs. Normally distributed ADC measurements were not significantly correlated with either total lung volume (r2 = 0.0001, P = 0.94) or menstrual age (r2 = 0.003, P = 0.70). The mean ADC value was 1.75 (95% CI, 1.63–1.86). Mean ± SD intraexaminer bias was −0.15 ± 2.3 (95% LOA, −4.7 to + 4.4) and interexaminer bias was 2.2 ± 3.5 (95% LOA, −4.7 to + 9.1).

Conclusions

Our findings suggest that ADC measurements of the fetal lung are reproducible between blinded examiners and are independent of menstrual age, as well as lung volume. Copyright © 2009 ISUOG. Published by John Wiley & Sons, Ltd.


Introduction

  1. Top of page
  2. Abstract
  3. Introduction
  4. Methods
  5. Results
  6. Discussion
  7. Acknowledgements
  8. REFERENCES

Fetal magnetic resonance imaging (MRI) is useful for the prenatal detection of selected congenital anomalies because it provides images with excellent tissue contrast resolution1–3. The MRI technology offers potential advantages over obstetric ultrasonography because a wider field of view is obtained without the usual acoustic shadowing that is often encountered during pregnancy. Diffusion-weighted imaging (DWI) provides another type of image contrast that is based on the molecular motion of water—a process that can be altered by the presence of disease4, 5. The apparent diffusion coefficient (ADC) is based on these principles and reflects water movement within the tissue environment. In this context, ADC values have been well described for non-obstetric applications, including the evaluation of newborns6, 7, adult brain ischemia8 and tumors9.

Only three other published studies have investigated the use of DWI to characterize water diffusion in human fetal lung tissue. In 2001, Moore et al. first introduced this concept with a very low static magnetic field strength of only 0.5 Tesla (T) to demonstrate that the ADC increased with gestational age10. Two other studies, however, subsequently reported conflicting results regarding the relationship between ADC and gestational age under a higher magnetic field strength of 1.5 T11, 12. The present paper describes our experience with DWI during the second and third trimesters of pregnancy. We document the range of ADC signal intensities for the fetal lung, reproducibility of ADC measurements, and the relationship of this parameter to menstrual age.

Methods

  1. Top of page
  2. Abstract
  3. Introduction
  4. Methods
  5. Results
  6. Discussion
  7. Acknowledgements
  8. REFERENCES

Research subjects were invited to participate under informed consent that was approved by the Human Investigation Committee at William Beaumont Hospital and the Institutional Review Board at the Eunice Shriver Kennedy National Institutes of Child Health and Human Development. All subjects had a previously normal prenatal ultrasound screening examination upon entry into this protocol. Pregnant volunteers were scanned primarily in a supine position on a 1.5-T MRI system (Sonata, Siemens Medical Solutions, Malvern, PA, USA). No maternal or fetal sedation was used. The MRI system was equipped with 40-mT/m gradients with a rise time of 0.2 ms (slew rate 200 T/m/s). A six-channel body array coil was positioned near the fetus for the optimal signal reception.

For conventional imaging, non-breath-hold HASTE and TRUE-FISP sequences provided proper localization and selection of a plane before implementing a rapid echo-planar sequence with DWI. Each DWI sequence took approximately 2 min. Single-shot DWI utilized TE/TR 125/3400 ms, slice thickness 4 mm, field of view 280 mm × 280 mm, interslice gap 0.8 mm and a matrix of 128 × 128. The voxel size was 2.5 mm × 2.5 mm × 4.0 mm. Two b-values (0 and 1000) were chosen along three orthogonal directions. Fat suppression was used with a pixel bandwidth of 1500 Hz. With a parallel imaging factor (GRAPPA) of 2, the scan time with three averages was reduced to 63 s. ADC maps were created using assigned b-values.

The highest-quality parallel slice acquisition series was chosen that utilized ADC maps of the fetal lung for each subject. Offline MRI software was used to analyze signal intensities from traced regions of interest (OsiriX DICOM software, http://www.osirix-viewer.com, WACOM Intuos graphics tablet). Total lung volume was calculated by manually tracing the best defined T2-weighted images for a given MRI acquisition sequence (Figure 1). Next, the traced diffusion maps were used to estimate the coefficient of variation in the ADC measurement between slices for each subject. This information was used to decide whether or not it was really necessary to average the ADC values from all parallel image slices or to more simply use the largest image slice as a representative sample of the lung ADC value. If only the largest image slice was necessary, then the mean ADC value from this representative slice would be used to compare the right vs. left lung as well as the relationship between ADC and menstrual age or total lung volume.

thumbnail image

Figure 1. Apparent diffusion coefficient map of the axial fetal chest at 35 weeks' menstrual age. The right lung has been traced manually to estimate the mean signal intensity.

Download figure to PowerPoint

The Shapiro–Wilks normality test was used to evaluate the distribution of measurements. Simple linear regression was performed with Pearson correlation coefficient. Measurement bias and agreement were examined with their 95% limits of agreement (LOA) using Bland–Altman plots13. A paired t-test was used to evaluate possible differences in ADC measurements between left and right lungs. Statistical significance was taken at P < 0.05.

Results

  1. Top of page
  2. Abstract
  3. Introduction
  4. Methods
  5. Results
  6. Discussion
  7. Acknowledgements
  8. REFERENCES

Forty-seven MRI examinations were performed from January 2006 to September 2008 at a mean ± SD menstrual age of 29.2 ± 4.5 (range, 20.9–37.1) weeks. Newborn infants were delivered at 39.2 ± 1.5 weeks with a birth weight of 3277 ± 455 g. All infants had a normal newborn physical examination.

Each subject was examined for consistency of ADC measurements across fetal lung image slices. A median of 5 (interquartile range (IQR), 4–7) slices were acquired for each subject and the median coefficient of variation for ADC measurements among slices was 5.6% (IQR, 4.0–8.1%). As the coefficient of variation was small, subsequent analysis of ADC values was based on the single largest lung slice from each fetus that was scanned using DWI. The mean ADC value was 1.75 (95% CI, 1.63–1.86).

Total lung volume significantly increased with menstrual age (r2 = 0.57, P < 0.001). However, normally distributed ADC measurements were not significantly correlated with either total lung volume (r2 = 0.0001, P = 0.94) or menstrual age (r2 = 0.003, P = 0.70) (Figures 2 and 3). The mean ± SD right lung ADC value was not statistically different from the mean left lung measurements (1.75 ± 0.40 vs. 1.71 ± 0.32; P = 0.10). Hence, subsequent analysis was based on ADC values from the right lung as it was easier to perform tracings without considering the fetal heart on the left side.

thumbnail image

Figure 2. Relationship of apparent diffusion coefficient (ADC) in fetal lung tissue to total tung volume. y = 0.0002x + 1.7; r2 = 0.0001, P = 0.94.

Download figure to PowerPoint

thumbnail image

Figure 3. Relationship of apparent diffusion coefficient (ADC) in fetal lung tissue to menstrual age. y = −0.05x + 1.9; r2 = 0.003, P = 0.70.

Download figure to PowerPoint

Reproducibility studies were performed for ADC measurements on the right lung. The mean ± SD intraexaminer bias and agreement was −0.1 ± 2.3% (95% LOA, −4.7 to 4.4%) (Figure 4) and the interexaminer bias and agreement was 2.2 ± 3.5% (95% LOA, −4.7 to 9.1%) (Figure 5).

thumbnail image

Figure 4. Bland–Altman plot of intraexaminer measurement bias with 95% limits of agreement for fetal lung apparent diffusion coefficient (ADC) measurements.

Download figure to PowerPoint

thumbnail image

Figure 5. Bland–Altman plot of interexaminer measurement bias with 95% limits of agreement for fetal lung apparent diffusion coefficient (ADC) measurements.

Download figure to PowerPoint

Discussion

  1. Top of page
  2. Abstract
  3. Introduction
  4. Methods
  5. Results
  6. Discussion
  7. Acknowledgements
  8. REFERENCES

Simple diffusion can be described as the random movement of molecules from areas of high to low concentration within a medium. The process also causes signal loss during conventional T1- and T2-weighted MRI sequences, although such effects are insignificant unless the relative signal change is amplified14. This is achieved by applying a strong magnetic field gradient as the movement of water molecules attenuates the observed signal within a magnetic field gradient. The DWI methodology applies a T2-weighted sequence with two extra gradient pulses with equal amplitude and width. They are placed on either side of a 180° radiofrequency pulse. Static proton spins are dephased by the first pulse and then completely rephased by a second pulse. If the net protons have been moved by diffusion along that axis during the time interval between pulses, they will not experience the same amount of phase shift during the presence of a magnetic field gradient. As a result, the returning signal will be attenuated. The degree of signal attenuation is proportional to the strength of the diffusion weighting which is denoted by a ‘b-value’, a factor that is dependent on gradient length, amplitude and the spacing between them.

The diffusion of protons in free water is actually greater than what occurs in biological tissue. Such a process is influenced by the presence of natural barriers to translational motion, such as cell membranes and myelin fibers (i.e. anisotropy). The ADC calculation in biological tissue accounts for quantitative diffusion that varies from the orientation of restricting barriers in relation to the diffusion-sensitive gradient direction. At least two b-values are required to estimate the water ADC value in a specific tissue of interest. If the b-value is too low, the diffusion-induced signal attenuation will be comparable to variance of the diffusion-weighted data and this result will underestimate the ADC value. When an excessively large b-value is chosen, the high signal attenuation and signal intensity may drop below the system noise level. The b-value is commonly set to 1000 s/mm2 in clinical practice and allows for sensitivity to changes in the diffusion coefficient while retaining an adequate signal-to-noise ratio15. As a final technical consideration, changes in lung perfusion can theoretically affect the ADC value and mask the effects of water diffusion. However, the echo-planar nature of our rapid measurement sequence (spin echo with echo-planar read-out) provides temporal resolution that is sufficient to minimize any effects from flow, perfusion and gross body movement.

Relatively few studies have examined the application of DWI for ADC measurements in fetal lung. Moore et al. were the first to use a low field strength MRI system (0.5 T) for establishing a normal range of fetal ADC values in 26 subjects10. Their results suggested a trend for ADC to increase with gestational age at a rate of 0.07 × 10−9 m2/s per week. A second study by Balassy et al. found small regional ADC differences within the right fetal lung that ranged from 1.91 to 2.13 µm2/ms when using a stronger 1.5-T MRI system11. Similar to the present study, ADC values did not significantly correlate with either menstrual age or lung volume. They proposed that the ‘relatively stable relation of the future air-space volume and the interstitium’ was the reason why ADC values did not correlate with menstrual age. A third investigation by Manganaro et al. also used a 1.5-T MRI system, on 50 subjects between 18 and 36 weeks' gestation12. Thirty-five of their fetuses had non-pulmonary anomalies. Their ADC values ranged from 1.2 µm2/ms at 18 weeks' gestation to 3.9 µm2/ms at 36 weeks' gestation, with a significant correlation between ADC values and age (r = 0.82, P < 0.001).

Some of these discrepancies may be related to methodological differences in the manner in which ADC measurements were obtained and compared with gestational age. For example, Moore et al. used a very low magnetic field strength gradient and globally averaged all ADC measurements from both lungs as they appeared on various images10. Manganaro et al. averaged ADC values from ‘all the lung parenchyma in all axial sections’12. Our results were most consistent with the findings of Balassy et al., who used the ‘best motion-artifact-free coronal slice that displayed the large surface of the lungs’ to estimate ADC of the right fetal lung11. Our study analyzed the largest image slice, regardless of the scanning plane, that provided good-quality data that could be used as a representative sample of fetal lung tissue.

The reproducibility of these measurements has not been extensively examined for fetal lung tissue. Moore et al. did not specifically address this issue using a lower static magnetic field strength10. Manganaro and et al. also did not examine the reproducibility of DWI measurements12. Balassy et al. reported that only two of the 30 average ADC measurements from three different lung regions exceeded the 95% LOA, although these measurements were obtained from only 10 subjects and gestational ages were not provided11. Similar to their results, we used the largest lung image slice for the ADC calculation. Our technique for averaging ADC values from a single image slice of the right lung simplified the alternative requirement for multiple manual tracings, and was associated with acceptable bias and agreement between examiners.

Validation of a reproducible ADC measurement technique for the fetal lung may provide a potential non-invasive biomarker of fetal lung maturity that is independent of both menstrual age and lung volume. Moore and colleagues investigated the in-vitro effects of varying surfactant concentrations on T1- and T2-weighted imaging relaxation times as well as ADC values10. None of these parameters, however, was significantly correlated with surfactant solution concentrations. They also examined a three-compartment in-vitro model that included intralung amniotic fluid, intratissue water and vascular blood. The introduction of a vascular component to the lung model was associated with a dramatic increase in ADC. The model predicted the in-vivo data reasonably well and they proposed that fetal lung diffusion measurements could represent a novel marker for development of capillaries surrounding the terminal lung tubules. They also proposed that ADC measurements could be used to study fetal lung maturation10. Although Balassy et al. concluded that ADC was not a good indicator of lung maturity, their study design did not specifically examine outcome variables that could be used to directly test this hypothesis11. Manganaro et al. also hypothesized that lung ADC measurements could be used to study fetal pulmonary maturity because of the correlation that was found between ADC and menstrual age12.

DWI can be used to characterize movement of water molecules through fetal tissue. Other investigators have described the use of this technology to evaluate the fetal brain16–19, kidneys20, 21 and twin–twin transfusion22. Our findings indicate that ADC measurements of the fetal lung are reproducible between blinded examiners and are independent of menstrual age. A simplified approach for tracing a well defined section of right lung appears to satisfactorily provide a representative sample of ADC values in the normal fetus. Future studies will need to correlate ADC measurements with amniotic fluid pulmonary surfactant and neonatal outcome to further examine the potential use of this non-invasive parameter for predicting the risk of respiratory distress syndrome.

Acknowledgements

  1. Top of page
  2. Abstract
  3. Introduction
  4. Methods
  5. Results
  6. Discussion
  7. Acknowledgements
  8. REFERENCES

The authors wish to acknowledge the technical assistance of Melissa Powell, RDMS, and Beverley McNie, BS, CCRP. This research was supported (in part) by the Perinatology Research Branch, Division of Intramural Research, Eunice Kennedy Shriver National Institute of Child Health and Human Development, NIH, DHHS.

REFERENCES

  1. Top of page
  2. Abstract
  3. Introduction
  4. Methods
  5. Results
  6. Discussion
  7. Acknowledgements
  8. REFERENCES