A randomised comparison of InnoScope and Macintosh laryngoscope in simulated difficult tracheal intubation in manikins

Authors

  • A. W. Y. Fong,

  • K. C. Lam,

  • B. C. P. Cheng,

  • K. K. Lam,

  • M. T. V. Chan

Errata

This article corrects:

  1. A randomised comparison of InnoScope and Macintosh laryngoscope in simulated difficult tracheal intubation in manikins Volume 68, Issue 2, 167–174, Article first published online: 15 November 2012

In the Summary of this paper [1], the sentence:

In this scenario, the median (interquartile range [range]) duration of tracheal intubation was significantly longer with the InnoScope (53 (20–100 [15–120]) s) compared with the Macintosh laryngoscope (27 (20–62 [15–120]) s), p = 0.01.

should have read:

In this scenario, the median (interquartile range [range]) duration for tracheal intubation was significantly longer with InnoScope compared with the Macintosh laryngoscope, 70 [19-120 (15-120)] s vs 30 [21–58 (15–120)] s, p = 0.01.

Table 1 was incorrectly reproduced and is printed here in its correct form.

Table 1. Conditions of laryngoscopy and tracheal intubation using InnoScope or Macintosh laryngoscope on a SimMan simulator with normal airway, difficult airway and limited neck movement
 InnoScopeMacintosh Laryngoscopep value
  1. POGO = Percentage of glottic opening. Values are number or median [Interquartile range (range)].

Normal Airway
 POGO scale; %100 [90–100 (70–100)]80 [65–100 (30–100)]0.003
 Duration of tracheal intubation; s 20 [14–25 (13–50)]18 [14–21 (10–32)]0.11
 Number of intubation attempts   1.00
1 2425 
2 1 0 
3 0 0 
Failed intubation 0 0 
Difficult Airway
 POGO scale; % 60 [30–90 (0–100)]40 [10–55 (0–95)]0.04
 Duration of tracheal intubation; s 70 [19–120 (15–120)]30 [21–58 (15–120)]0.01
 Number of intubation attempts   0.002
1 1224 
2 1 0 
3 4 0 
 Failed intubation 8 1 
Limited Neck Movement
 POGO scale; % 90 [80–100 (30–100)]50 [30–75 (0–100)]< 0.001
 Duration of tracheal intubation; s 16 [13–26 (10–120)]16 [14–25 (10–32)]0.14
 Number of intubation attempts   0.20
1 1225 
2 2 0 
3 0 0 
Failed intubation 1 0 

The authors apologise for these errors.