Clinical relevance of CD157 for rapid and cost-effective simultaneous evaluation of PNH granulocytes and monocytes by flow cytometry

Authors


Summary

Introduction

CD157 has been reported as a potentially useful marker for paroxysmal nocturnal hemoglobinuria (PNH) testing by flow cytometry (FCM).

Methods

We determined the performance characteristics of a CD157-based five-color assay and compared results from patient analysis with results obtained with a previously validated CD14/CD24-based six-color protocol.

Results

Coefficient of variation (CV) for intra-/interassay precision analysis of granulocytes ranged from 0.88/0.09% to 3.02/3.71% and from 0.20/0.08% to 8.83/4.04% for the five- and six-color protocol, respectively. For monocyte, CV ranged from 0.42/0.49% to 8.13/4.80% for the five-color protocol and from 0.28/0.70% to 5.41/3.19% for the six-color protocol within various PNH clones. Coefficient of determination (r2) for linear regression analysis of PNH clones ranging from 0.3 to 99.8% was >0.99 in all cases, Wilcoxon ranks test showed no statistically significant differences (> 0.05), and Bland–Altman analysis demonstrated agreement with mean bias ranging from −0.02 to 0.38.

Conclusion

Our results confirm very good performance characteristics for both intra- and interassay precision analyses, excellent correlation, and agreement between approaches. In agreement with recently published data, our experience supports the clinical relevance of CD157 for a rapid and cost-effective simultaneous evaluation of PNH leukocytes by flow cytometry.