Checking Anaesthetic Equipment 2012

Association of Anaesthetists of Great Britain and Ireland

Authors


  • This is a consensus document produced by expert members of a Working Party established by the Association of Anaesthetists of Great Britain and Ireland (AAGBI). It has been seen and approved by the AAGBI Council.

  • This article is accompanied by an Editorial. See page 571 of this issue.

  • Reuse of this article is permitted in accordance with the Creative Commons Deed, Attribution 2.5, which does not permit commercial exploitation.

  • You can respond to this article at http://www.anaesthesiacorrespondence.com

Summary

A pre-use check to ensure the correct functioning of anaesthetic equipment is essential to patient safety. The anaesthetist has a primary responsibility to understand the function of the anaesthetic equipment and to check it before use. Anaesthetists must not use equipment unless they have been trained to use it and are competent to do so. A self-inflating bag must be immediately available in any location where anaesthesia may be given. A two-bag test should be performed after the breathing system, vaporisers and ventilator have been checked individually. A record should be kept with the anaesthetic machine that these checks have been done. The ‘first user’ check after servicing is especially important and must be recorded.