Background  Colonic propagating sequences (PS) are important for the movement of colonic content and defecation, and aberrant PS patterning has been associated with slow transit constipation. However, because these motor patterns are typically recorded over long periods (24 h +), the visualization of PS spatiotemporal patterning is difficult. Here, we develop a novel method for displaying pan-colonic motility patterns.

Methods  A 3D mesh representing the geometry of the human colon was created as follows: (i) Human colon images from the Visible Human Dataset were digitized to create a 3D data cloud, and (ii) A surface mesh was fitted to the cloud using a least-squares minimization technique. Colonic manometry catheters were placed in the ascending colon of healthy controls and patients with slow transit constipation (STC), with the aid of a colonoscope. The colonic manometry data were interpolated and mapped to the model according to the following anatomical landmarks: cecum, hepatic flexure, splenic flexure, sigmoid-descending junction, and anus.

Key Results  These 3D images clearly and intuitively communicate characteristics of normal and abnormal colonic motility. Specifically we have shown the reduced amplitude of the antegrade propagating pressure waves (PPW) throughout the colon and reduced frequency of PPWs at the mid-colon in patients with STC.

Conclusions and Inferences  A novel method for the 3D visualization of PS is presented, providing an intuitive method for representing a large volume of physiological data. These techniques can be used to display frequency, amplitude or velocity data, and will help to convey regions of abnormally in patient populations.


Introduction

  1. Top of page
  2. Abstract
  3. Introduction
  4. Methods
  5. Results
  6. Discussion
  7. Acknowledgments
  8. Author contribution
  9. Competing interest
  10. References

Constipation is a common and distressing condition that conveys a major socioeconomic and disease burden.1,2 Dysregulated colonic contractile activity is implicated in constipation3,4 yet quantitative data unequivocally differentiating normal from abnormal colonic function is lacking.5 This is due to several technical and analytical barriers that prevent the accurate manometric profiling of colonic luminal pressure patterns.

Previous colonic manometric studies have utilized catheters with ≤16 sensors spaced at ≥5-cm intervals.6,7 While these studies have provided important insights into dysmotility associated with constipation, they lack the detail currently achieved by high-resolution (HR) manometry approaches in other gut regions, notably the esophagus.8 To overcome this problem, we recently developed a novel high-resolution fiber-optic manometry catheter capable of pan-colonic recordings at 1-cm intervals.9

Whilst HR colonic manometry offers the potential to yield substantial new insights into normal and abnormal colonic motility, it also brings major analytical challenges. Colonic manometry recordings are generally taken over hours or days, generating a vast data set. Key data features must be extracted and distilled in an efficient analysis pipeline, and communicated via intuitive visualization methods that allow for rapid identification of general or regional abnormalities.4 Moreover, whilst we have previously developed a technique of spatiotemporal mapping colonic propagating pressure waves,10 that technique is limited to displaying data from a single subject. Displaying grouped mean data of pressure wave regional frequency or amplitude still relies on graphs, which become excessively complex with HR recordings.

This study introduces novel solutions for current analysis needs in both low and HR colonic manometry, enabling the anatomical registration and visualization of colonic manometry data, via three-dimensional (3D) virtual modeling.

Methods

  1. Top of page
  2. Abstract
  3. Introduction
  4. Methods
  5. Results
  6. Discussion
  7. Acknowledgments
  8. Author contribution
  9. Competing interest
  10. References

Patients and controls

Presented data were obtained from two published sources: (i) 24-h pan-colonic water perfused manometry in eight healthy controls and 14 patients with slow transit constipation (STC) (16 sensors at 7.5-cm resolution),4 and (ii) 24-h fiber-optic high resolution pan-colonic recordings in a patient with STC (72 sensors at 1-cm resolution).9 Methods and protocols for water perfused and fiber-optic manometric recordings have been described in detail elsewhere.4,9

Data extraction

Propagating pressure waves (PPWs) were identified via semi-automated marking using custom software (St. George Clinical School, Sydney, Australia),written in Matlab (The MathWorks, MA, USA.) and Java (Sun Microsystems, CA, USA.). Definitions of propagating sequences (PSs) and PPWs have been previously defined;11 briefly, a PS was defined as an array of three or more propagating pressure waves recorded from adjacent recording sites in which the conduction velocity between wave onsets within that sequence lay between 0.2 and 12 cm/s.11 Propagating sequences were further qualified by the terms antegrade or retrograde, depending upon the direction of propagation.

Anatomical registration and visualization

The 3D model framework employed in this study was translated from methods described by Fernandez et al.12 A 3D data cloud was created by digitizing the surface of the human colon using serial slices of the Visible Human Dataset.13 A 3D finite element surface was then fitted to the data cloud, using a least squares minimization technique.12 Fixed anatomical reference points were identified at the cecum, hepatic and splenic flexures, the descending-sigmoid colon junction, and the anus.

For the water perfused studies, the colon was divided into 16 regions (region 1, cecum; region 4, hepatic flexure; region 8, splenic flexure; region 12, proximal sigmoid colon; region 16, rectum).11 Recording side holes were assigned to the colonic region within which they lay, with side hole one always in region 1.4 For the HR studies, a plain abdominal X-ray was taken (Fig. 1G), and all recording sensors were identified and assigned anatomical sites according to their observed locations.

image

Figure 1.  (A and B) displays the interpolation method used to map the discrete pressure transducer data to the 3D virtual colon. (C–F) show pressure amplitude (mmHg) and PPW frequency (cumulative frequency) from data in Fig. 2C and D. (C and E) show data from the control groups. (D and F) represent the patient groups. The color spectrum denotes low activity/pressure with white, and high/activity pressure with dark blue/red, respectively. Note that in health, the amplitude of PPWs increase at the splenic flexure and descending colon (C), and that this region also represents the area of greater wave frequency (E). In contrast, the amplitude of PPWs remains constantly low throughout the colon in patients with slow transit constipation (D) and a reduced frequency of PPWs occurs at the distal transverse colon and splenic flexure (F). A fluoroscopic image of the HR catheter in a subsection of the colon is shown in (G) with the cumulative frequency of PPWs waves recorded in this individual displayed in (H). A paucity of activity is evident in the splenic flexure and descending colon.

Download figure to PowerPoint

Manometry data from the above anatomical sites were translated to the corresponding points on the geometric mesh. The remaining data between each of the fixed neighboring points were linearly interpolated along the center line of the virtual colon. This data manipulation was achieved using scripts written in the PERL programing language,a which allowed the manometry data to be easily converted to a file format that could be read into the visualization component of the cmiss software,b called cmgui visualization software. CMISS is a mathematical modeling and simulation environment for solving bioengineering problems developed at the University of Auckland. CMGUI was then applied to display the data in the visual format (i.e. spectrum range, orientation etc.) desired.c This technique insured that the displayed pressure at anatomical points of interest was preserved when transferred from the patient data to the model. The interpolated pressures were then projected from the center line of the colon onto the 3D walls of the virtual colon. For example, Fig. 1A shows a theoretical set of pressure transducers (large spheres), colored with a pressure distribution, located on the virtual catheter. Figure 1B shows the data linearly interpolated along the virtual catheter. The interpolated data are then projected orthogonally onto the walls of the virtual colon to produce images such as Figs 1C–F, H.

To validate these methods, we present the cumulative frequency and amplitude of the antegrade PPWs per colonic region. These data have been recently been published in a less intuitive format.4,9 A white–blue spectrum was used to visualize the PPW cumulative frequency data over the 3D model, where white indicates no pressure waves and dark blue indicates the highest frequency of PPWs. The average amplitudes of PPWs per colonic regions are shown using a white–red spectrum.

Results are displayed together with previous methods (spatiotemporal maps and standard histograms) for comparison.4

Results

  1. Top of page
  2. Abstract
  3. Introduction
  4. Methods
  5. Results
  6. Discussion
  7. Acknowledgments
  8. Author contribution
  9. Competing interest
  10. References

Standard spatiotemporal maps of a healthy control and a patient with STC are presented in Fig. 2A and B. These demonstrate the visualization of propagating sequence data for individual subjects, highlighting abnormal motility profiles.4 However, grouped mean data cannot be displayed by these methods, and are instead presented as histograms (Fig. 2C and D). These histograms can be unintuitive with multi-sensor data and become prohibitively complex with HR data (72 sensors). Histograms could be simplified by binning data into anatomical regions (e.g. ascending, transverse, and descending colon), however that would defeat the purpose of recording with multiple sensors.

image

Figure 2.  (A and B) show spatiotemporal maps of propagating activity in a single healthy control (A) and a slow transit constipation patient (B).4 The ridges represent antegrade (green) and retrograde (red) propagating sequences (PS). The maps demonstrate the spatiotemporal distribution and amplitude of PSs. Histograms are used to display the amplitude (C) and wave frequency (D) data over the colonic regions for groups of subjects.

Download figure to PowerPoint

Grouped mean data presented using the novel 3D modeling approaches are shown in Figs 1C–F. These images clearly and intuitively communicate characteristics previously described in STC patients: the reduced amplitude of the antegrade PPWs (Fig. 1D) and the paucity of PPWs at the distal transverse and splenic flexure (Fig. 1F).4

Importantly, the novel techniques developed here are ideally suited to HR data sets from individual or grouped patients. Figure 1G shows the fiber-optic catheter positioned in a STC patient. The cumulative frequency of the PPWs recorded at each sensor over a 24-h period is shown interpolated on Fig. 1H. In this patient, PPWs are shown to occur in the transverse and sigmoid colon, but there is a marked reduction of activity throughout the entire descending colon.

Discussion

  1. Top of page
  2. Abstract
  3. Introduction
  4. Methods
  5. Results
  6. Discussion
  7. Acknowledgments
  8. Author contribution
  9. Competing interest
  10. References

Novel approaches for the anatomical registration and 3D visualization of pan-colonic manometry data are presented. This ‘pipeline’ of methods effectively condenses a vast volume of complex manometric data into an efficient and intuitive display format, while allowing the data to be interpreted according to its original context. Furthermore, these strategies allow the presentation of group mean pressure data with substantially greater clarity than is possible with histograms.4 These methods are shown to be ideally suited to new HR colonic manometric approaches.9

Most significantly, these methods permit all observers to easily distinguish the normal physiology of PPW frequency and amplitude from the pathophysiological characteristics in patients with dysmotility.

The pathophysiology of constipation remains poorly understood. An ongoing lack of objective colonic pressure measurements in adults continues to limit clinical advances, with only two small interventional studies being based on colonic manometry evidence.3,14 In contrast, clinicians now routinely use esophageal manometry to subtype swallowing disorders and plan treatment.8 There are several reasons for this discrepancy. The esophagus is easily accessible and has relatively controlled motility. In addition, HR esophageal manometry is now widely practiced and spatiotemporal pressure profile plots have become standard tools, quickly allow clinicians to learn and apply the technique.15

Pan-colonic HR manometry is now possible, following the advent of fiber-optic pressure sensing technology,9 and together with these new anatomical registration and visualization techniques, we now have established improved potential to acquire, analyze and interpret colonic motility data, to identify manometric markers of dysmotility, and potentially to define new subtypes and mechanisms of constipation.

Important limitations to these methods should be noted. Foremost, applying an individual’s data to a generic colon negates the considerable variations in anatomy. This is partly overcome by mapping the data to fixed anatomical reference points. However, in future, it would be more desirable to develop a library of generic templates, from which an appropriate geometric match can be selected, as is standard practice in functional brain imaging.16 The geometries generated routinely using CT colonography, such as those presented in Hanson et al.17 would provide an excellent source of patient specific or library models. Finally, the interpolated grouped mean data do not allow for this display of the SEM or the SD at each colonic region. However, the main purpose of methods is to accurately portray manometric data traits, and variance can easily be presented in accompanying tables or text.

Footnotes

Acknowledgments

  1. Top of page
  2. Abstract
  3. Introduction
  4. Methods
  5. Results
  6. Discussion
  7. Acknowledgments
  8. Author contribution
  9. Competing interest
  10. References

The authors are supported by grants from the NZ Health Research Council, Riddet Institute, the NHMRC Australia, and NIH (R01 DK64775).

Author contribution

  1. Top of page
  2. Abstract
  3. Introduction
  4. Methods
  5. Results
  6. Discussion
  7. Acknowledgments
  8. Author contribution
  9. Competing interest
  10. References

The novel display method was developed by JB Davidson, PG Dinning and G O’Grady; the high resolution catheter was developed by JW Arkwright; data was obtained by PG Dinning; AJ Pullan contributed technical oversight. Full access to data was provided. All authors read and approved the final manuscript. This work was presented in abstract form at the 2010 NGM meeting in Boston.18