Optical spectroscopy techniques can accurately distinguish benign and malignant renal tumours

Authors


Abstract

What's known on the subject? and What does the study add?

  • There is little known about optical spectroscopy techniques ability to evaluate renal tumours.
  • This study shows for the first time the ability of Raman and optical reflectance spectroscopy to distinguish benign and malignant renal tumours in an ex vivo environment. We plan to develop this optical assistance in the operating room in the near future.

Objective

  • To evaluate the ability of Raman spectroscopy (RS) and optical reflectance spectroscopy (ORS) to distinguish benign and malignant renal tumours at surgery.

Materials and Methods

  • Between March and October 2011, RS and ORS spectra were prospectively acquired on surgical renal specimens removed for suspicion of renal cell carcinoma (RCC). Optical measurements were done immediately after surgery.
  • Optical signals were normalised to ensure comparison between spectra. Initial and final portions of each spectrum were removed to avoid artefacts.
  • A support vector machine (SVM) was built and tested using a leave-one-out cross-validation.
  • Classification scores, including accuracy, sensitivity and specificity were calculated on the entire population and in patients with tumours of <4 cm.

Results

  • In all, 60 nephrectomies were performed for 53 malignant tumours (41 clear-cell, eight papillary and four chromophobe carcinomas) and seven benign tumours (four oncocytomas, two angiomyolipomas and a haemorrhagic cyst). In all, >700 optical spectra were obtained and submitted to SVM classification.
  • The SVM could recognise benign and malignant renal tumours with an accuracy of 96% (RS) and 88% (ORS) in the whole population and with an accuracy of 93% (RS) and 95% (ORS) in the present subset of small renal tumours (<4 cm).
  • Histological subtype could be determined in 80% and 88% of the cases with RS and ORS, respectively.
  • The SVM was able to differentiate chromophobe carcinomas and benign lesions with an accuracy of 96% in RS and 98% in ORS.

Conclusion

  • Benign and malignant renal tumours can be accurately discriminated by a combination of RS and ORS. In vivo experiments are needed to further assess the value of optical spectroscopy techniques.
Abbreviations
ORS

optical reflectance spectroscopy

OS

optical spectroscopy

RCC

renal cell carcinoma

SVM

support vector machine

RS

Raman spectroscopy

Introduction

RCC is the third most common urological malignancy after prostate and bladder cancer [1]. Because of the widespread use of abdominal imaging, its incidence has increased over the last 20 years. Still, the natural evolution of RCC is sometimes difficult to predict, and up to 25% of patients with a localised tumour at diagnosis will develop metastases [2]. Various biomarkers have been suggested to have a prognostic value to foretell the course of RCC but to date none is used in clinical practice. It is therefore interesting to consider new approaches to RCC.

Optical spectroscopy (OS) is a group of novel technologies that are based on the interaction of light with tissue. Raman spectroscopy (RS) and optical reflectance spectroscopy (ORS) are the most commonly used OS techniques. RS uses a laser beam and ORS uses a simple white-light source. OS provides optical spectra that give information on the composition of the tissue and can therefore be described as an ‘optical biopsy tool’. It has been proven that OS could precisely differentiate histological subtype and/or grade of various solid tumours including prostate [3], bladder [4] and kidney [5-8] cancers.

In the present study, we prospectively evaluated the ability of RS and ORS to differentiate benign and malignant renal tumours as well as histological subtypes among RCCs.

Materials and Methods

Our method of RS acquisition has been described previously [5, 9]. Briefly, the RS system comprised three major elements: (i) A high powered near-infrared laser (Invictus NIR laser 785 nm, 500 mW, Kaiser Optical Systems, Ann Arbor, MI, USA), (ii) A fibre optic magnetic resonance probe, (iii) A RXN1 Raman analyser (Kaiser Optical Systems), and (iv) A laptop computer with HoloGRAMS© 4.1 software for Raman spectra acquisition.

The ORS system included: (i) a deuterium tungsten light source emitting light in the wavelength range of 200 to 2000 nm, (ii) a fibre optic array probe, (iii) a 200 to 1100 nm USB2000 spectrophotometer (Ocean Optics, Dunedin, FL, USA), (iv) A laptop computer with SpectraSuite© software (Ocean Optics).

After approval by our Institutional Review Board, we prospectively collected RS and ORS spectra in consecutive kidney specimens removed due to the suspicion of cancer. Between March and October 2011, 64 operations were performed, including 37 radical nephrectomies and 27 partial nephrectomies. Four patients were excluded based on final histology results: one with liposarcoma, one with an Xp11 translocation RCC, one with a mixed epithelial and stromal tumour, and one with a totally necrotic tumour. Surgery was completed through an open access in 16 patients and laparoscopically (with or without robotic assistance) in 44 patients. Immediately after extraction, the specimens were stored on ice and transferred to the pathology department. Spectra were acquired ≤15 min of extraction.

For RS, we used the same parameters as in our previous experiments [5, 9]: laser beam was set at 50 mW and focused on a 500 μm spot on the surface of the sample. Acquisition period was 30 s. A wavelength of 785 nm was chosen to minimise background autofluorescence due to natural fluorophores.

For ORS, before each session of measurements, we acquired a calibration spectrum from a standard surface. This standard surface provides ≈100% of reflectance in the wavelength range of 250–1000 nm. We also recorded a background measurement with the same experimental setting but with the light source turned off. This background data was subtracted from each optical measurement during data analysis. The acquisition period was 10 s.

RS and ORS spectra were acquired in a complete dark environment to avoid artefacts due to ambient light (Fig. 1). After longitudinal sectioning of the specimens bythe pathologist, areas without necrosis or blood clots were selected for optical acquisition. Spectra were obtained in several locations inside and on the surface of normal and tumour tissue. We obtained a total of 787 spectra (419 RS and 368 ORS).

Figure 1.

OS acquisition device: tumoral kidney is placed under the light source and the spectra are acquired with the black box completely sealed to avoid signal contamination by ambient light.

We used the same statistical method for RS and ORS spectra. Analysis of the signal comprised four steps:

  1. Normalisation of the signal by dividing every spectral point by the area under the curve to ensure all spectra were comparable. All spectra were of optimal quality.
  2. Initial (λ < 150 cm−1 in RS and λ < 250 nm in ORS) and final (λ > 830 nm in ORS) portions of the spectra were removed to avoid artefacts. Samples of benign and malignant spectra are shown in Figure 2.
  3. Our objective was to establish a classification tool that could recognize (i) benign and malignant tumours, (ii) histological subtype of malignant tumours (i.e. clear cell, papillary and chromophobe), and (iii) chromophobe and benign tumours. For that purpose, we built a support vector machine (SVM). A SVM is a mathematical tool that allows selecting relevant features for classifying a dataset into two groups [10]. After training, the SVM is tested with a leave-one-out cross-validation technique (Fig. 3): for every classification, all spectra but one are used to train the SVM and the remaining spectrum is set aside for final testing [8]. The process was repeated with alternation of the testing spectrum for all spectra. At the end of the test, a cross-validation score that represents the ability of a single spectrum to classify the assessed biological sample was obtained. We trained the SVM in the entire population and in the subset of patients with small renal tumours (<4 cm), as it is in this group that being able to diagnose a benign tumour is most useful in clinical practice.
  4. A uniform and independent random integer number generator function was used in association with the SVM when there was a discrepancy between the sample numbers of spectra. For example, we had many more malignant than benign spectra (193 vs 33, respectively); randomisation process allowed the comparison of an equal number of 33 benign and malignant spectra. The process was repeated 10 times. The risk of over- or under-training of the SVM was thus limited.
Figure 2.

Raman spectra of benign (blue) and malignant tumoral (red) renal tissue.

Figure 3.

SVM illustration.

Results

The clinicopathological characteristics of the population are given in Tables 1 and 2. Histology was malignant in 53 cases (41 clear-cell, eight papillary, and four chromophobe tumours), and benign in seven cases (four oncocytomas, two angiomyolipomas, and a haemorrhagic cyst).

Table 1. Patient and tumour characteristics
VariableValue
No. of patients60
Mean (sd): 
Age, years61.3 (12.3)
Tumour size, cm6.1 (3.7)
N: 
Type of surgery 
Radical nephrectomy: 
Open12
Laparoscopic21
Partial nephrectomy: 
Open4
Robotic23
Pathology: 
Benign7
Malignant53
Histological subtype of malignant tumours:
Clear-cell carcinoma41
Papillary8
Chromophobe4
Tumour stage: 
pT1a19
pT1b4
pT25
pT323
pT42
Nodal invasion: 
Nx-052
N11
Distant metastases: 
M046
M17
Tumour grade: 
Low22
High31
Table 2. Patient and tumour characteristics for small renal masses (<4 cm)
VariableValue
No. of specimens28
Mean (sd): 
Age, years60.5 (12)
Tumour size, cm3 (0.9)
N: 
Type of surgery: 
Radical nephrectomy: 
Open1
Laparoscopic5
Partial nephrectomy: 
Open3
Robotic19
Pathology: 
Benign4
Malignant24
Histological subtype of malignant tumours:
Clear-cell carcinoma16
Papillary5
Chromophobe3
Tumour stage: 
pT1a20
pT1b1
pT20
pT33
Tumour grade: 
Low15
High9

Classification scores are presented in Tables 3 and 4. Overall, the SVM could differentiate benign from malignant renal tumours with an accuracy of 96% (RS) and 88% (ORS). In the subset of malignant tumours, the SVM was able to correctly recognise clear-cell carcinomas 88% and 89% of the times with RS and ORS, respectively. The SVM could set apart chromophobe carcinomas and benign lesions with an accuracy of 96% (RS) and 98.4% (ORS).

Table 3. RS and ORS classification scores for the whole population
 RSORS
Accuracy, %Sensitivity, %Specificity, %Accuracy, %Sensitivity, %Specificity, %
Benign vs malignant96.599.193.987.895.680
Clear-cell vs others88.386.490.289.184.194.1
Papillary vs others80.685.975.288.291.285.2
Chromophobe vs others92.995.390.689.797.981.6
Benign vs chromophobe95.993.598.298.496.8100
Table 4. RS and ORS classification scores in small renal masses (<4 cm)
 RSORS
Accuracy, %Sensitivity, %Specificity, %Accuracy, %Sensitivity, %Specificity, %
Benign vs malignant93.197.288.99597.992.1
Clear-cell vs others85.48585.794.792.696.8
Papillary vs others77.18074.186.791.781.7
Chromophobe vs others92.494.590.488.896.181.5
Benign vs chromophobe87.385.489.199.699.3100

When focusing on patients with small renal tumours, the ability of the SVM to differentiate benign from malignant lesions was 93% and 95% for RS and ORS, respectively. Clear-cell carcinomas could be identified with 85% and 94.7% accuracies with RS and ORS, respectively. Chromophobe carcinomas and benign tumours could be discriminated by the SVM in 87% and 99.6% of the cases, with RS and ORS respectively.

Discussion

OS techniques rely on the interaction of light with tissues. RS is based on the inelastic scattering of light: when photons hit a tissue, their energy raises the molecules' vibrational state. When a molecule returns to its background level, one photon out of 108 is emitted at a different wavelength from the incident light. This change in wavelength is known as the ‘Raman shift’. ORS is based on differences of absorption of light depending on its wavelength. The analysis of RS and ORS spectra provides information on the composition and structure of a tissue. RS and ORS showed promising results for ex vivo evaluation of several solid tumours [11-15]. More recently, spectroscopic techniques have been used to assess urological tumours [3-8, 16-22]. For example, it has been shown that RS could accurately identify bladder tumour stage and grade [4, 19, 21, 22], or differentiate inflammatory, benign and malignant prostatic tissue [3, 22].

In the field of RCC, we previously showed that RS could accurately differentiate tumoral from normal renal tissue as well as histological subtype and tumour grade [5]. However, our number of benign spectra was too limited for the algorithm to recognise benign and malignant renal tumours. The present study, based on a significantly larger population, showed that both RS and ORS could efficiently differentiate benign from malignant renal tumours with an accuracy >90%. The sensitivity and specificity of our model were optimal (99.1% and 95.6% for RS and ORS, respectively). The performances of RS and ORS remained excellent in the subset of small renal tumours (<4 cm) with precision rates of 93% and 95%, respectively. This is particularly interesting as in clinical practice, the most challenging situation when one needs to know whether a renal tumour is benign or malignant, is in the context of small renal masses. RS and ORS could also effectively recognise histological subtypes of malignant tumours with precision rates ranging between 80% and 93%.

We specifically focused on the ability of OS techniques to discern chromophobe tumours from benign lesions. Chromophobe carcinoma accounts for 5% of malignant RCC and is generally thought to have a rather indolent profile compared with clear-cell tumours [23]. Overlapping morphological characteristics often make it difficult to differentiate chromophobe carcinoma from renal oncocytoma [24]. Moreover chromophobe carcinoma can coexists with oncocytoma inside of a hybrid tumour [25]. Diagnosis of a hybrid tumour can be challenging particularly when made on a percutaneous biopsy [26, 27]. The present results showed that the SVM could precisely differentiate chromophobe carcinomas from benign lesions with an accuracy >95%.

The data concerning OS is rapidly growing. However, reports on urological tumours are still limited [3-5, 8, 19, 21, 22]. There has been an extensive research in oncology to better appraise the natural course of cancer. Many biomarkers in the blood, genes and elsewhere have been investigated [28], but none has proven enough reliability and reproducibility to be used in clinical practice. In this situation, OS is a promising technique with many potential advantages: (i) It gives immediate results (ii) It is an objective method using robust algorithms that does not need human interpretation, i(ii) It is not invasive, and (iv) It is not toxic and can easily be transferred to human care.

There have been several publications that specifically addressed the interest of OS to evaluate RCC. The team from the University of Texas showed that ORS could accurately discriminate normal and tumoral renal tissue at surgery [7]. We previously reported in a series of 36 patients that RS could differentiate normal and tumoral renal tissue, high- and low-grade tumours, as well as histological subtype of RCC [5, 8]. In the present study, we focused on optical differences of benign and malignant renal tumours. It comprises more patients than our previous experiment with a significant number of benign lesions (seven of 60). Moreover, we refined our technique of optical acquisition by using a completely sealed black box to avoid artefacts due to ambient light.

We used two OS techniques (RS and ORS) that complement each other. RS requires heavy equipment but provides spectra with very rich information. Each molecule has its own pattern of vibration that can serve as a ‘Raman biomarker’. RS would therefore be a promising method to assist the pathologist for improving diagnostic accuracy. It is well known, for example, that the establishment of Fuhrman grade is subject to intra- and inter-observer variability [29]. There have been recent reports of computational programs that could outperform human pathologists in distinguishing between grades of breast cancers and provide a more accurate prognosis than a human pathologist could manage [30]. Conversely, ORS is a simpler, cheaper and less bulky system that would be ideal for use in the operating room to help the surgeon in real-time to differentiate between a malignant and benign tumour.

The present study has several limitations. Our population, although extended, remains quite small. Optical measurements were acquired in an ex vivo environment, so the impact of blood flow has not been taken into account. Haemoglobin is the preponderant chromophore and has a high capacity to absorb light. For this reason there is a need for in vivo studies. For now, both Raman and reflectance probes are not approved for human use. We are therefore working on a specific thin probe that could be directly applied on the tumour surface (in case of open surgery) or inserted through a trocar (during laparoscopy) and provide optical information in real-time. We also plan to make optical measurements of frozen tissues stored in a tumour database to delineate optical features predictors of survival. Finally, comparing traditional histological interpretation to optical measurements would also be an interesting issue. We are therefore planning to compare spectra acquired from renal tumour core biopsies with the ‘blinded’ interpretation of an experienced uropathologist to see whether the objective information provided by OS can help reduce the 10% uncertainty usually associated with renal biopsy analysis.

In conclusion, the present study showed that a combination of OS techniques with the help of appropriate signal processing approaches could accurately distinguish between benign and malignant renal tumours ex vivo. These results support recent publications to consider OS as a promising tool for the evaluation of kidney tumours.