SEARCH

SEARCH BY CITATION

Keywords:

  • BACTEC;
  • BacTALERT;
  • blood culture;
  • direct diagnostic;
  • MALDI-TOF MS

Clin Microbiol Infect 2011; 17: 192–195

Abstract

Recently, matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) has been presented as a novel method for the direct identification of bacteria from positive blood culture bottles. The rate of the MALDI TOF MS-based identification in the present study from positive BacT/ALERT (bioMérieux, Marcy l’Etoile, France) blood culture bottles was 30%, which is far below the previously reported sensitivities using the BACTEC (Becton Dickinson, Franklin Lakes, NJ, USA) system. We also found evidence that the Biotyper algorithm did not identify a second pathogen in cases of positive BacT/ALERT blood culture bottles containing two different species.