Validity of Transdermal Alcohol Monitoring: Fixed and Self-Regulated Dosing
Article first published online: 6 JAN 2006
DOI: 10.1111/j.1530-0277.2006.00004.x
Additional Information
How to Cite
Sakai, J. T., Mikulich-Gilbertson, S. K., Long, R. J. and Crowley, T. J. (2006), Validity of Transdermal Alcohol Monitoring: Fixed and Self-Regulated Dosing. Alcoholism: Clinical and Experimental Research, 30: 26–33. doi: 10.1111/j.1530-0277.2006.00004.x
Publication History
- Issue published online: 6 JAN 2006
- Article first published online: 6 JAN 2006
- Received for publication May 4, 2005; accepted September 19, 2005.
- Abstract
- Article
- References
- Cited By
Keywords:
- Alcohol Monitoring;
- Transdermal;
- Insensible Perspiration;
- Sweat
Background: To study the validity of transdermal assessment of alcohol concentration measured by a lightweight, noninvasive device.
Methods: Subjects wore a 227-g anklet that sensed transdermal alcohol concentrations (TACs) every 15 to 30 minutes, downloading results to a remote computer each day. Twenty-four subjects entered a laboratory and received a dose of 0, 0.28, or 0.56 g/kg of ethanol. Breath alcohol concentrations (BrAC) and TAC were measured every 15 to 30 minutes Twenty others [10 alcohol dependent (AD) and 10 not (NAD)] in the community who wore the anklet for 8 days kept a drinking log and provided a BrAC sample each day.
Results: In the laboratory, no zero-dose subject, and every subject receiving alcohol, had alcohol-positive TACs. The device distinguished low- and high-alcohol–dosing groups using peak (t14=3.37; p<0.01) and area under the curve (t14=3.42; p<0.01) of TACs. Within dosing groups, average TAC curves were broader (right-shifted) and had lower peaks than average BrAC curves. For community participants, self-reported number of drinks (t18=−3.77; p<0.01), area under the TAC curve (t9.5=−3.56; p<0.01), and mean TAC (t9.9=−3.35; p<0.01) all significantly distinguished the AD and NAD groups. However, individual transdermal readings were not reliably quantitatively equivalent to simultaneously obtained breath results.
Conclusions: Within the limits of the laboratory study, the device consistently detected consumption of approximately 2 standard drinks. On average, the device shows discriminative validity as a semiquantitative measure of alcohol consumption but individual readings often are not equivalent to simultaneous BrACs.