Misidentification of maternal heart rate as fetal on cardiotocography during the second stage of labor: the role of the fetal electrocardiograph

Authors


  • Conflict of interest Except the first author, all authors are involved in teaching midwives and obstetricians in intrapartum fetal monitoring in several centers in the United Kingdom and abroad. The authors have stated explicitly that there are no conflicts of interest in connection with this article.

  • Please cite this article as: Nurani R, Chandraharan E, Lowe V, Ugwumadu A, Arulkumaran S. Misidentification of maternal heart rate as fetal on cardiotocography during the second stage of labor: the role of the fetal electrocardiograph. Acta Obstet Gynecol Scand. 2012;91: DOI: 10.1111/j.1600-0412.2012.01511.x

Edwin Chandraharan, MBBS, MS (Obs & Gyn), DFFP, MRCOG, St. George's Healthcare NHS Trust, Blackshaw Road, London SW 17 0RE, UK. E-mail: edwin.c@sky.com

Abstract

Objective: To identify the incidence of fetal heart rate (FHR) accelerations in the second stage of labor and the role of fetal electrocardiograph (ECG) in avoiding misidentification of maternal heart rate (MHR) as FHR. Design: Retrospective observational study. Setting: University hospital labor ward, London, UK. Sample: Cardiotocograph (CTG) tracings of 100 fetuses monitored using external transducers and internal scalp electrodes. Methods: CTG traces that fulfilled inclusion criteria were selected from an electronic FHR monitoring database. Main outcome measures: Rate of accelerations during external and internal monitoring as well as decelerations for a period of 60 minutes prior to delivery were determined. The role of fetal ECG in differentiating between MHR and FHR trace was explored. Results: Decelerations occurred in 89% of CTG traces during the second stage of labor. Accelerations indicating possible recording of FHR or MHR were found in 28.1 and 10.9% of cases recorded by an external ultrasound transducer as well as internal scalp electrode, respectively. Accelerations coinciding with uterine contractions occurred only in 11.7 and 4% of external and internal recording of FHR, respectively. Absence of ‘p-wave’ of the ECG waveform was associated with MHR trace. Conclusion: Decelerations were the commonest CTG feature during the second stage of labor. The incidence of accelerations coinciding with uterine contractions was less than half in fetuses monitored using a fetal scalp electrode. Analysing the ECG waveform for the absence of ‘p-wave’ helps in differentiating MHR from FHR.

Get access to the full text of this article

Ancillary