High Definition Oscillometry: a novel technique for non-invasive blood pressure monitoring in the cynomolgus monkey (Macaca fascicularis)

Authors


Dr med. vet. Barthel Schmelting, PhD, Covance Laboratories GmbH, Kesselfeld 29, 48163 Muenster, Germany.
Tel.: +49 251 97 98 0;
fax.: +49 251 78 46 97;
e-mail: barthel.schmelting@covance.com

Abstract

Background  Current approaches for accurate blood pressure determination rely predominantly on invasive techniques. High Definition Oscillometry (HDO) was evaluated as a potential non-invasive approach for accurate blood pressure recordings in cynomolgus monkeys.

Methods  In conscious animals, systolic, diastolic, mean arterial blood pressure (MAP) and pulse/minute were determined 15 times within approx. 9 minutes per individual. This session was performed during 3 consecutive days. Anaesthesia induced hypotension was controlled simultaneously with HDO and telemetry as reference.

Results  Repeated measurements were highly reproducible. After procedural habituation, mean MAP was 96.2 ± 13.7 mmHg in males and 86.9 ± 4.3 mmHg in females. Mean intraindividual coefficients of variation ranged between 10.8% and 2.4% depending on the session and parameter. Values determined by HDO corresponded to those reported for invasive techniques.

Conclusion  Our results demonstrate, using telemetry as reference, the accuracy of HDO-based non-invasive blood pressure measurements in macaques to detect drug-related cardiovascular changes.