A Prospective Study Comparing the Sensed R Wave in Bipolar and Extended Bipolar Configurations: The PropR Study

Authors

  • ANEESH V. TOLAT M.D.,

    Corresponding author
    • Saint Francis Hospital and Medical Center, Section of Electrophysiology, University of Connecticut School of Medicine, Hartford, Connecticut
    Search for more papers by this author
  • MELISSA WOICIECHOWSKI M.S.N.,

    1. Saint Francis Hospital and Medical Center, Section of Electrophysiology, University of Connecticut School of Medicine, Hartford, Connecticut
    Search for more papers by this author
  • ROSEMARIE KAHR R.C.I.S.,

    1. Saint Francis Hospital and Medical Center, Section of Electrophysiology, University of Connecticut School of Medicine, Hartford, Connecticut
    Search for more papers by this author
  • JOSEPH DELL'ORFANO M.D.,

    1. Saint Francis Hospital and Medical Center, Section of Electrophysiology, University of Connecticut School of Medicine, Hartford, Connecticut
    Search for more papers by this author
  • ELLISON BERNS M.D.,

    1. Saint Francis Hospital and Medical Center, Section of Electrophysiology, University of Connecticut School of Medicine, Hartford, Connecticut
    Search for more papers by this author
  • BRUCE BERNSTEIN Ph.D.,

    1. St. Christopher's Hospital for Children, Section of General Pediatrics, Drexel University College of Medicine, Philadelphia, Pennsylvania
    Search for more papers by this author
  • NEAL LIPPMAN M.D.

    1. Saint Francis Hospital and Medical Center, Section of Electrophysiology, University of Connecticut School of Medicine, Hartford, Connecticut
    Search for more papers by this author

  • Dr. Lippman has reported being a speaker for Medtronic. None of the other authors have reported any disclosures relevant to this study.

  • This study was supported in part by an investigator initiated grant from Medtronic.

Address for reprints: Aneesh V. Tolat, M.D., F.A.C.C., F.H.R.S., 1000 Asylum Ave., Suite 3206, Hartford, CT 06105. Fax: 860 714 9993; e-mail: atolat@ctheartbeat.com

Abstract

Background

Progress in implantable cardiac defibrillator (ICD) technology has allowed for switching the sensing polarity for the detection of ventricular fibrillation (VF). However, whether one sensing polarity confers additional advantage over the other is not known.

Objectives

To determine whether one sensing polarity is superior to the other for the detection of VF.

Methods

Patients were enrolled into a prospective randomized study of sensing of VF and R waves in normal rhythm. Sensing of VF was determined by number of under sensed beats (USB), and time to detection of VF (TDVF). Each patient underwent ICD implantation followed by testing of the ICD. At each induction, patients were randomized to sensing in extended bipolar (EBP) or true bipolar (TBP) configuration. Additionally, R waves were compared at implant and at 1-month follow-up.

Results

A total of 50 patients were enrolled into the study. When evaluating the primary endpoint, no difference was found between USB in EBP or TBP configuration; 1.1 ± 1.2 beats versus 1.3 ± 1.3 beats; P = NS. Also, no difference was found between TDVF in EBP or TBP configurations; 5.9 ± 0.6 seconds versus 5.9 ± 0.6 seconds; P = NS. With regard to the secondary endpoints, there was no difference between R waves in EBP or TBP configurations at the time of implant 10.9 ± 4.8 mV versus 10.9 ± 4.8 mV P = NS; or at 1-month follow-up 12.4 ± 4.7 mV versus 12.0 ± 5.4 mV P = NS.

Conclusions

There is no difference in the detection of VF between EBP or TBP configurations in patients undergoing ICD implantation.