Central corneal thickness measurements in normal dogs: a comparison between ultrasound pachymetry and optical coherence tomography

Authors


  • Funding: None; Optovue® iVue SD-OCT on loan from Optovue Inc. Freemont, CA.

Abstract

Objective

To compare central corneal thickness (CCT) measurements obtained from normal canines using a spectral-domain optical coherence tomography device (SD-OCT) and an ultrasonic pachymeter.

Animals

Fifteen clinically normal dogs.

Procedures

Dogs were briefly anesthetized with propofol. Central corneal thickness measurements were obtained from both eyes of each animal four times by one operator using a SD-OCT system. The CCT was automatically calculated using the pachymetry software generated from eight radial scans, 6 mm in length. Ultrasonic pachymetry was then performed in both eyes of each animal by one operator five times in succession. The CCT was recorded for each modality and compared statistically for agreement and bias.

Results

Fifteen dogs (seven castrated male, eight spayed female) with a mean age of 2.3 ± 1.3 years were examined. The mean CCT for all eyes examined via SD-OCT and ultrasonic pachymetry (velocity set at 1636 m/s) was 587.72 ± 32.44 μm and 598.54 ± 32.28 μm, respectively. Ultrasonic pachymetry consistently overestimated CCT as compared to OCT by a mean value of 10.82 μm (P < 0.0001). No significant difference in CCT was found based on the age or sex of the animal. There was no significant difference in CCT measurements between replicates performed by the same operator for each modality. The coefficient of variation within and between imaging modalities was low (range 0.63–2.5%). The intraclass correlation coefficient comparing methods was 0.92.

Conclusions

Correlation between modalities was excellent; however, ultrasonic pachymetry set at a standard velocity of 1636 m/s overestimates CCT as compared to OCT.