Influence of field size on a PTW type 23342 plane-parallel ionization chamber's response

Influence of field size on PTW 23342 chamber

Authors


Abstract

The response of a PTW type 23342 plane-parallel ionization chamber, both in air and in phantom, was evaluated for x-ray tube potentials between 30 and 100kV and radiation field diameters ranging from 30 to 70mm. The experiments were performed with a calibrated Pantak x-ray machine and made use of the same set of x-ray qualities adopted by the PTB primary laboratory for the calibration of such chambers. A Plexiglas® phantom (1.18gcm3) 110mm long, 110mm wide, and 80mm deep was used for phantom measurements. X-ray qualities were characterized by using 99.99% pure aluminum filters. On the basis of the IAEA's TRS 398, the article discusses the dependence of the plane-parallel ionization chamber readings with field size in air and in phantom, its implication with regard to clinical dosimetry, cross-calibration, and dissemination of calibration factors.