SU-E-J-110: TG 51 Dosimetry : With Or Without Lead

Authors


Abstract

TG-51 Dosimetry: With or Without Lead.

Purpose:

In this project, an analytical method has been introduced for adjustment of the TG-51 recommended KQ in order to produce accurate dosimetric data for high energy photons without the lead foil.

Methods:

These investigations were performed using a 30 cm × 30 cm × 30 cm CIVCO water tank, A12 EXRADIN Water proof Farmer Chamber, a Standard Imaging MAX 4000 electrometer, and 1 mm thick lead foil from Standard Imaging. Complete TG-51 was performed every month with and without lead. The results were analyzed and an analytical model has been developed for comparing the values of KQ. TG-51 Table I was used to obtain KQ values.

Results:

The dosimetric evaluations were obtained for Varian Linear accelerators Model 21ix and 21ex. These results indicates that the measured data with lead foil in place as recommended by TG-51 is in excellent agreement (within 0.1%) with the calculated data obtained by the new model, from our dosimetry data without-lead. If equation 15 of the TG-51 report is used without any adjustments, it will lead to differences of about 1.6 % (on the average) in relative data which will Resultin differences of about 0.3 % (on the average) in the KQ Values. The KQ value for 18 MV obtained consistently with the equation of TG-51 “with lead” and “without lead” were 0.971 and 0.974, respectively. The 0.3 % higher results for KQ without lead eventually will lead to 0.3% larger output. However, by considering this model the KQ value was found to be 0.971 for dosimetry without lead.

Conclusion:

The analytical model that was introduced in this project was able to reproduce the dosimetric data of the high energy linear accelerators to within 0.1% without the use of the lead foil.

Ancillary