SU-E-T-464: On the Equivalence of the Quality Correction Factor for Pencil Beam Scanning Proton Therapy

Authors

  • Sorriaux J,

    1. Center of Molecular Imaging, Radiotherapy and Oncology, Institut de recherche experimentale et Clinique, Universite catholique de Louvain, Avenue Hippocrate 54, 1200 Brussels, Belgium
    2. ICTEAM Institute, Universite catholique de Louvain, 1348 Louvain-la-Neuve, Belgium
    3. Departement of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA 02114, USA
    Search for more papers by this author
  • Paganetti H,

    1. Departement of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA 02114, USA
    Search for more papers by this author
  • Testa M,

    1. Departement of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA 02114, USA
    Search for more papers by this author
  • Giantsoudi D,

    1. Departement of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA 02114, USA
    Search for more papers by this author
  • Schuemann J,

    1. Departement of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA 02114, USA
    Search for more papers by this author
  • Bertrand D,

    1. Ion Beam Applications S.A, Louvain-la-Neuve, Belgium
    Search for more papers by this author
  • Orban de Xivry J.,

    1. ICTEAM Institute, Universite catholique de Louvain, 1348 Louvain-la-Neuve, Belgium
    Search for more papers by this author
  • Lee J,

    1. Center of Molecular Imaging, Radiotherapy and Oncology, Institut de recherche experimentale et Clinique, Universite catholique de Louvain, Avenue Hippocrate 54, 1200 Brussels, Belgium
    2. ICTEAM Institute, Universite catholique de Louvain, 1348 Louvain-la-Neuve, Belgium
    Search for more papers by this author
  • Palmans H,

    1. EBG MedAustron GmbH, Wiener Neustadt, Austria
    2. National Physical Laboratory, Teddington, UK
    Search for more papers by this author
  • Vynckier S,

    1. Departement de radiotherapie, Cliniques Universitaires Saint-Luc, Avenue Hippocrate 54, 1200 Brussels, Belgium
    Search for more papers by this author
  • Sterpin E

    1. Center of Molecular Imaging, Radiotherapy and Oncology, Institut de recherche experimentale et Clinique, Universite catholique de Louvain, Avenue Hippocrate 54, 1200 Brussels, Belgium
    Search for more papers by this author

Abstract

Purpose:

In current practice, most proton therapy centers apply IAEA TRS-398 reference dosimetry protocol. Quality correction factors (kQ) take into account in the dose determination process the differences in beam qualities used for calibration unit and for treatment unit. These quality correction factors are valid for specific reference conditions. TRS-398 reference conditions should be achievable in both scattered proton beams (i.e. DS) and scanned proton beams (i.e. PBS). However, it is not a priori clear if TRS-398 kQ data, which are based on Monte Carlo (MC) calculations in scattered beams, can be used for scanned beams. Using TOPAS-Geant4 MC simulations, the study aims to determine whether broad beam quality correction factors calculated in TRS-398 can be directly applied to PBS delivery modality.

Methods:

As reference conditions, we consider a 10×10×10 cm3 homogeneous dose distribution delivered by PBS system in a water phantom (32/10 cm range/modulation) and an air cavity placed at the center of the spread-out-Bragg-peak. In order to isolate beam differences, a hypothetical broad beam is simulated. This hypothetical beam reproduces exactly the same range modulation, and uses the same energy layers than the PBS field. Ion chamber responses are computed for the PBS and hypothetical beams and then compared.

Results:

For an air cavity of 2×2×0.2 cm3, the ratio of ion chamber responses for the PBS and hypothetical beam qualities is 0.9991 ± 0.0016.

Conclusion:

Quality correction factors are insensitive to the delivery pattern of the beam (broad beam or PBS), as long as similar dose distributions are achieved. This investigation, for an air cavity, suggests that broad beam quality correction factors published in TRS-398 can be applied for scanned beams.