Liver 4DMRI: A retrospective image-based sorting method

Authors

  • Paganelli Chiara,

    1. Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Milano 20133, Italy
    Search for more papers by this author
  • Summers Paul,

    1. Division of Radiology, Istituto Europeo di Oncologia, Milano 20133, Italy
    Search for more papers by this author
  • Bellomi Massimo,

    1. Division of Radiology, Istituto Europeo di Oncologia, Milano 20133, Italy and Department of Health Sciences, Università di Milano, Milano 20133, Italy
    Search for more papers by this author
  • Baroni Guido,

    1. Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Milano 20133, Italy and Bioengineering Unit, Centro Nazionale di Adroterapia Oncologica, Pavia 27100, Italy
    Search for more papers by this author
  • Riboldi Marco

    1. Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, Milano 20133, Italy and Bioengineering Unit, Centro Nazionale di Adroterapia Oncologica, Pavia 27100, Italy
    Search for more papers by this author

Abstract

Purpose:

Four-dimensional magnetic resonance imaging (4DMRI) is an emerging technique in radiotherapy treatment planning for organ motion quantification. In this paper, the authors present a novel 4DMRI retrospective image-based sorting method, providing reduced motion artifacts than using a standard monodimensional external respiratory surrogate.

Methods:

Serial interleaved 2D multislice MRI data were acquired from 24 liver cases (6 volunteers + 18 patients) to test the proposed 4DMRI sorting. Image similarity based on mutual information was applied to automatically identify a stable reference phase and sort the image sequence retrospectively, without the use of additional image or surrogate data to describe breathing motion.

Results:

The image-based 4DMRI provided a smoother liver profile than that obtained from standard resorting based on an external surrogate. Reduced motion artifacts were observed in image-based 4DMRI datasets with a fitting error of the liver profile measuring 1.2 ± 0.9 mm (median ± interquartile range) vs 2.1 ± 1.7 mm of the standard method.

Conclusions:

The authors present a novel methodology to derive a patient-specific 4DMRI model to describe organ motion due to breathing, with improved image quality in 4D reconstruction.