TU-H-CAMPUS-IeP1-02: Validation of a CT Monte Carlo Software

Authors


Abstract

Purpose:

To validate the in-house developed CT Monte Carlo calculation tool GMctdospp against reference simulation sets provided by the AAPM in the new report 195.

Methods:

Deposited energy was calculated in four segments (test 1) and two 10 cm long cylinders (test 2) inside a CTDI phantom (following case #4 of the AAPM report 195). The x-ray point source of a given 120 kVp spectrum was collimated to a fan beam with two thicknesses (10 mm, 80 mm) for a static and a rotational setup. In addition, a given chest geometry was used to calculate deposited energy in several organs for a 0° static and a rotational beam (following case #5 of the AAPM report 195). The results of GMctdospp were compared against the particular mean value of the four quoted Monte Carlo codes (EGSnrc, Geant 4, MCNP and Penelope).

Results:

Calculated values showed no outliers in any of the cases. Differences between GMctdospp and the particular mean

Results:

Calculated values showed no outliers in any of the cases. Differences between GMctdospp and the particular mean value were always at similar magnitude compared to the quoted codes. For case #4 (CTDI phantom) the relative differences were within 1.5 %, on average 0.4 % and for case #5 (chest phantom) within 2.5 % and on average 0.85 %.

Conclusion:

The results confirmed an overall uncertainty of the Monte-Carlo calculation chain in GMctdospp being <2.5 %, for most cases even better. This can be considered small compared to other sources of uncertainties, e.g. virtual source and patient models. The photon transport implemented in GMctdospp inside a voxel-based patient geometry was successfully verified.