Performance of a clinical gridded electron gun in magnetic fields: Implications for MRI-linac therapy

Authors

  • Whelan Brendan,

    1. Radiation Physics Laboratory, University of Sydney, Sydney, NSW 2006, Australia; Ingham Institute for Applied Medical Research, Liverpool, NSW 2170, Australia; and Liverpool Cancer Therapy Centre, Liverpool Hospital, Liverpool, NSW 2170, Australia
    Search for more papers by this author
  • Holloway Lois,

    1. South Western Clinical School, University of New South Wales, Sydney, NSW 2170, Australia; Institute of Medical Physics, School of Physics, University of Sydney, NSW 2006, Australia; and Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW 2522, Australia
    Search for more papers by this author
  • Constantin Dragos,

    1. Radiological Science Laboratory, Stanford University, Palo Alto, California 94305
    Search for more papers by this author
  • Oborn Brad,

    1. Illawarra Cancer Care Centre, Wollongong, NSW 2500, Australia and Centre for Medical Radiation Physics, University of Wollongong, Wollongong, NSW 2522, Australia
    Search for more papers by this author
  • Bazalova-Carter Magdalena,

    1. Department of Physics and Astronomy, University of Victoria, Victoria, British Columbia V8P 5C2, Canada
    Search for more papers by this author
  • Fahrig Rebecca,

    1. Radiological Science Laboratory, Stanford University, Palo Alto, California 94305
    Search for more papers by this author
  • Keall Paul

    1. Radiation Physics Laboratory, University of Sydney, Sydney, NSW 2006, Australia and Ingham Institute for Applied Medical Research, Liverpool, NSW 2170, Australia
    Search for more papers by this author

Abstract

Purpose

MRI-linac therapy is a rapidly growing field, and requires that conventional linear accelerators are operated with the fringe field of MRI magnets. One of the most sensitive accelerator components is the electron gun, which serves as the source of the beam. The purpose of this work was to develop a validated finite element model (FEM) model of a clinical triode (or gridded) electron gun, based on accurate geometric and electrical measurements, and to characterize the performance of this gun in magnetic fields.

Methods

The geometry of a Varian electron gun was measured using 3D laser scanning and digital calipers. The electric potentials and emission current of these guns were measured directly from six dose matched true beam linacs for the 6X, 10X, and 15X modes of operation. Based on these measurements, a finite element model (FEM) of the gun was developed using the commercial software opera/scala. The performance of the FEM model in magnetic fields was characterized using parallel fields ranging from 0 to 200 G in the in-line direction, and 0–35 G in the perpendicular direction.

Results

The FEM model matched the average measured emission current to within 5% across all three modes of operation. Different high voltage settings are used for the different modes; the 6X, 10X, and 15X modes have an average high voltage setting of 15, 10, and 11 kV. Due to these differences, different operating modes show different sensitivities in magnetic fields. For in line fields, the first current loss occurs at 40, 20, and 30 G for each mode. This is a much greater sensitivity than has previously been observed. For perpendicular fields, first beam loss occurred at 8, 5, and 5 G and total beam loss at 27, 22, and 20 G.

Conclusions

A validated FEM model of a clinical triode electron gun has been developed based on accurate geometric and electrical measurements. Three different operating modes were simulated, with a maximum mean error of 5%. This gun shows greater sensitivity to in-line magnetic fields than previously presented models, and different operating modes show different sensitivity.