Gradient echo imaging

Michael Markl PhD,

Corresponding Author

Department of Radiology, Feinberg School of Medicine, Northwestern University, Chicago, Illinois, USA

Department of Biomedical Engineering, McCormick School of Engineering, Northwestern University, Evanston, Illinois, USA

Departments of Radiology and Biomedical Engineering, Northwestern University, 737 N. Michigan Ave., Ste. 1600, Chicago, IL 60611Search for more papers by this author
Jochen Leupold PhD,

Department of Radiology, Medical Physics, University Medical Center Freiburg, Freiburg, Germany

Search for more papers by this author
First published: 15 May 2012
Citations: 35

Abstract

Magnetic resonance imaging (MRI) based on gradient echoes is used in a wide variety of imaging techniques and clinical applications. Gradient echo sequences form the basis for an essential group of imaging methods that find widespread use in clinical practice, particularly when fast imaging is important, as for example in cardiac MRI or contrast-enhanced MR angiography. However, the term “gradient echo sequence” is somewhat unspecific, as even images acquired with the most common sequences employing the gradient echo for data acquisition can significantly differ in signal, contrast, artifact behavior, and sensitivity to, eg, flow. This is due to the different use of sequence timing and basic sequence building blocks such as spoiler gradients or specific radiofrequency (RF) pulse phase patterns. In this article the basic principles of gradient echo formation compared to spin echo imaging are reviewed and the properties of gradient echo imaging in its simplest form (TR ≫ T2) are described. Further, the most common three variants of fast gradient echo sequences (TR < T2), namely, unbalanced gradient echo, RF spoiled gradient echo, and balanced steady state free precession; are discussed. For each gradient echo sequence type, examples of applications exploiting the specific properties of the individual technique are presented. J. Magn. Reson. Imaging 2012;35:1274–1289. © 2012 Wiley Periodicals, Inc.

PRINCIPLES AND CONCEPTS of gradient echo (also “gradient recalled echo” [GRE]) imaging are the basis of many applications on modern magnetic resonance imaging (MRI) systems. Technically, the difference between spin echo (1-3) and gradient echo imaging is related to the pulse sequence elements that are used to generate an MR signal (Figure 1). While two radiofrequency (RF) pulses (90° and 180°) are used for spin refocusing and spin echo generation, GRE imaging is based on only a single RF pulse, typically <90°, in combination with readout gradient reversal (4-7). As a result, shorter repetition times (TR) and therefore faster imaging is feasible. Although signal formation is also more sensitive to magnetic field inhomogeneity caused by hardware-related imperfections of the main magnetic field (shim) or properties of tissue (magnetic susceptibility) (8-10), today's advanced hardware, particularly improved shim and gradient performance, permits the successful application of GRE imaging with reliable and reproducible image quality. Many GRE-based applications such as dynamic (CINE) cardiac imaging (11), MR angiography (12), assessment of tissue perfusion or viability (13-16), or body or musculoskeletal imaging (17, 18) are an integral part of MR protocols used in routine clinical settings.

image

Simplified spin echo (top) and gradient echo (bottom) pulse sequence diagrams. The basic difference between gradient echo and spin echo imaging is related to the fact that echo formation is a result of a single RF pulse and gradient reversal while spin echo imaging uses two RF pulses, ie, a second 180° pulse, for echo generation. As a result, the gradient echo signal intensity is determined by Turn:x-wiley:10531807:media:JMRI23638:tex2gif-stack-1-decay and, contrary to the spin echo T2-weighted signal intensity, field inhomogeneity effects are not refocused at the time of signal formation. DAQ reflects the period of data acquisition. See section Relaxation Effects for a more detailed description of T2 and Turn:x-wiley:10531807:media:JMRI23638:tex2gif-stack-2-decay.

The purpose of this review is to provide an overview of the fundamental properties of GRE imaging and to describe the influence of the selection of imaging parameters on signal, contrast, and typical artifacts. Next to the fundamental properties of gradient echo formation and their consequences regarding signal and contrast of basic GRE imaging (TR ≫ T2), the three most common fast GRE techniques (TR < T2) (19) are discussed in this review: balanced steady state free precession (bSSFP), unbalanced GRE, and RF spoiled gradient echo imaging. In addition, the effects of contrast agent, flow, magnetic susceptibility, and chemical shift are briefly reviewed and examples of the clinical application of different GRE techniques are provided. For further details, the reader may consult textbooks on MRI (20-22).

MR Spin Magnetization Dynamics

The MR signal is created by the nuclear proton spin magnetization of 1H hydrogen atoms (23). Inside an MR system the strong magnetic field B0 along the z-axis of the scanner (direction of the magnet bore) will result in the alignment of a certain fraction of the spin magnetization along the direction of B0. As a result, a net “thermal equilibrium” magnetization M directed along B0 is generated (Figure 2, left). The z-component of M characterizes the magnetization that is available for signal generation, which can be converted into transverse magnetization by RF excitation. The application of RF excitation with a flip angle α tips the available magnetization M toward the transverse plane by the angle α (Figure 2, mid). The resulting magnetization vector can be characterized by its remaining longitudinal component Mz along the direction of the main magnetic field and a transverse component Mxy in the plane orthogonal to the main magnetic field B0. Once tipped in the transverse plane, Mxy continuously rotates around the main magnetic field with the Larmor frequency ωL and induces a detectable MR signal, the so-called free induction decay (FID) in a receive coil surrounding the object under investigation. The speed of the rotation of Mxy (precession frequency ωL) is directly determined by the local magnetic field, ie, changes in the local magnetic field due to tissue properties (regional differences in magnetic susceptibility, eg, at air–tissue interfaces) or hardware imperfections (quality of the main magnetic field, shim) will result in locally different rotation frequencies.

image

Spin magnetization during thermal equilibrium (a), after RF excitation (b), and following the dephasing of the transverse magnetization Mxy (c).

Relaxation Effects

The MR signal will persist as long as Mxy has a nonzero component. Two different relaxation effects (T1-recovery and T*2-decay) will affect the amplitude of the signal. T1-recovery characterizes restoration of the longitudinal magnetization Mz, ie, the magnetization returns to its equilibrium magnetization M directed along B0 by exchanging energy with its environment. Typically, T1 relaxation values are field strength-dependent (longer T1 times at higher fields) and can vary between ≈250 msec (fat at 1.5 T), ≈700–800 msec (brain tissue at 1.5T), ≈1200 msec (blood at 1.5 T), and up to ≈3–4 sec (cerebrospinal fluid at 1.5 T).

T1 relaxation can thus reduce the transverse magnetization Mxy. More important, however, local magnetic field fluctuations and associated T*2-effects will result in dephasing of the transverse magnetization Mxy and a rapid signal decay on a much shorter time scale, within milliseconds to seconds (Figure 2, right). Based on local variations of the magnetic field due to a variety of effects (susceptibility differences at air–tissue interfaces, imperfect shim, interactions between magnetic moments of neighboring spins) the transverse magnetization Mxy experiences different magnetic fields locally. As a result, the transverse magnetization rotates at locally different frequencies and, in a given amount of time, magnetization at different locations acquires different amounts of rotation in the transverse plane (dephasing, Figure 3, right). The differences in amount of rotation occur on a small scale such that within each voxel of an MR image the net detectable transverse magnetization Mxy is reduced by this dephasing effect (the vector addition of the individual magnetization components with different amount of rotation results in a reduced net transverse magnetization which determines the detectable signal for each voxel).

image

Signal formation for spin echo (top) and gradient echo (bottom) imaging. Top: The 180° refocusing pulse for spin echo imaging flips over magnetization and reverses the rotation direction of the transverse magnetization and results in a compensation of dephasing caused by the T2′ effect resulting in the rephasing of the magnetization at echo time TE. The colored arrows represent exemplary magnetization vectors. Bottom: For gradient echo imaging, the dephasing is not reversed. Signal and contrast are determined by both T2′ and T2 effects. Note that shorter echo times are necessary for detectable gradient echo signal intensity. [Color figure can be viewed in the online issue, which is available at wileyonlinelibrary.com.]

The dephasing of the transverse magnetization is governed by two independent processes. The decay time T*2 is composed of spin dephasing caused by static (time-invariant) dephasing effects (T2′-decay) and nonconstant (time-variant) field fluctuations (T2-decay) as described by the following relationship:
equation image(1)
T2′-decay reflects the dephasing due to temporally unchanged magnetic field inhomogeneity. These originate from, eg, local differences of the magnetic properties (described by the so-called susceptibility) of tissues, fluids, contrast agents, etc., or from hardware-related magnetic field imperfections (eg, shim). In contrast, T2-decay is caused by stochastic and temporal varying interactions between the magnetic moments of neighboring spins. Due to the stochastic nature of this process, the dephasing effects associated with T2-decay cannot be reversed with a spin echo sequence. In contrast, dephasing caused by deterministic T2′ can be fully time-reversed with the spin-echo.

Spin and Gradient Echo Formation

For a spin echo experiment (Figs. 1, 3, top), two RF pulses (90° and 180°) are used for echo generation. During the time between the 90° and 180° pulses, the transverse magnetization Mxy starts to dephase according to T2′ and T2 effects. The second RF pulse with a flip angle of 180°, called a refocusing pulse, is used to flip over (invert) the magnetization which reverses the rotation direction of all transverse magnetization components Mxy. Dephasing that has occurred as result of static (T2′) effects, such as susceptibility or field imperfections (9), is reversed. This process leads to the formation of a spin echo at echo time TE when the T2′-dephasing has effectively been nullified. Spin echo imaging thus compensates for static magnetic field effects and the signal strength is governed by the remaining T2-induced dephasing and signal decay.

The situation is different for GRE imaging (Figs. 1, 3, bottom), which uses a single RF pulse in combination with readout gradient reversal such that the net gradient area is zero at echo time TE. The gradient echo sequence omits the formation of a spin echo and directly uses the signal from the free induction decay following the RF excitation pulse. Mechanisms such as field inhomogeneity or susceptibility are not refocused with the gradient echo at echo time TE and will influence the signal and contrast. Note that T*2 includes both T2 and static T2′-field effects (Figure 3) and is typically much shorter than T2 alone. As a result, GRE signal intensity decays much faster and echo times have to be shorter than for spin echo imaging in order to yield sufficient signal intensity.

In short, while spin echoes are RF-refocused and provide T2-dependent signal strength, GRE imaging is gradient refocused and has echo amplitudes determined by T*2-decay (Figure 3). Note that the formation of a (gradient) echo instead of a pure FID leads to better quality of the magnitude image. The symmetric shape of the echo is beneficial to the Fourier transform. Recording a pure FID (exponential decay) with a readout gradient only and without a preceding dephasing gradient would be detrimental to image quality (24).

Basic GRE Imaging (TR ≫ T2)

MRI typically consists of several repetitions of a basic pulse sequence module with given repetition and echo times TR and TE until all data for a complete 2D or 3D dataset are collected, as illustrated in Figure 4. As a consequence, the timing of the acquisition, ie, TE and TR relative to signal recovery (T1) and decay (T2, T*2), will determine signal intensity and contrast. A gradient echo sequence acts as a pure progressive saturation sequence on longitudinal magnetization Mz, such that a so-called steady state will build up after several RF excitations.

image

Steady state formation for basic GRE imaging (TR ≫ T2). Repeated RF excitation results in a reduction of the longitudinal magnetization (Mz) for consecutive repetition times (TR) until the steady state magnetization (MSS) is reached. Mxy fully decays prior to each RF-excitation pulse and does not contribute to the steady state formation. The steady state magnetization (MSS) which defines the signal intensity for basic GRE imaging is determined by T1, TR, and the flip angle α.

To illustrate such a steady state formation, Figure 4 shows the temporal evolution of the magnetization components Mz and Mxy over several pulse sequence repetitions for basic GRE imaging with repetition times TR much greater than T2. Due to the long TR, the transverse magnetization Mxy (dashed lines) completely decays prior to each new RF excitation. However, since TR is typically shorter than, or on the order of T1, the longitudinal magnetization (solid lines) will not fully recover to its initial amplitude M0. Instead, after several repetitions a new longitudinal equilibrium magnetization MSS is reached, the “steady state.” The signal intensity in the MR image is determined by the available longitudinal magnetization MSS in the steady state that can be converted into detectable MR signal. The steady state value MSS can be calculated according to the so-called “Ernst equation” (25):
equation image(2)
which determines the signal intensity for basic GRE imaging. Basic GRE images are thus T1-weighted and the signal intensity is determined by the ratio TR/T1 and the flip angle α. Signal intensity can be maximized for a given TR and T1 (ie, tissue) using an optimal flip angle α = arccos[exp(-TR/T1)], also known as the “Ernst angle.” In many imaging applications, optimizing the signal from one tissue is not the major concern. More important is optimizing the contrast between two tissues (eg, lesion and healthy tissue). It should be noted that even though the signal intensities may be optimal for certain Ernst angles, contrast between tissues with different T1 is typically maximized at higher flip angles (26, 27).

The steady state formation and relationship between T1, TR, and flip angle described above is the reason why flip angles much smaller than 90° are typically chosen for GRE imaging as compared to spin echo imaging for which maximum signal intensity is provide by 90°–180° RF pulse pairs.

Figure 5 summarizes the signal and contrast behavior for GRE imaging in the Ernst regime. Any T2 contributions are minimized and images are predominantly T1-weighted. The highest signal intensities are typically found at low flip angles (between 40° and 50°). Tissue contrast, however, is optimized for larger flip angles (for example, gray and white matter contrast is maximized using flip angles between 60° and 80°).

image

Cranial images acquired with basic GRE imaging (TR = 500 msec, TE = 4 msec, TR ≫ T2) demonstrate signal and contrast behavior as a function of flip angle. PD, proton density.[Color figure can be viewed in the online issue, which is available at wileyonlinelibrary.com.]

Practically, basic GRE imaging is often not very favorable because the need for a long TR leads to long acquisition times. However, the application of RF spoiling (see next section) offers the possibility of making the Ernst equation valid for short repetition times TR.

Fast GRE Imaging (TR < T2)

Fast GRE imaging refers to the situation where TR is assumed to be short (TR ≪ T1 and < T2), such that a steady state of the magnetization has built up during image acquisition (28-30). Since repetition times TR are short (<T2), the transverse magnetization Mxy does not fully decay prior to each new RF pulse and therefore contributes to the steady state formation. In contrast to basic GRE imaging, both the longitudinal (Mz) and the transverse (Mxy) magnetization determine the final steady state signal intensity. As for basic GRE, signal intensities demonstrate strong flip angle dependence but also more complex contrast behavior, depending not only on T1 TR, and but also T2 (29). For this review, three different types of fast GRE sequences are distinguished, depending on how the transverse magnetization is used during the sequence TR, as illustrated in Figure 6.

image

Comparison of the three variants of fast gradient echo (GRE) imaging. Note that all three images were acquired with short TE and TR. Unbalanced GRE (left) shows mixed T2/T1 contrast and cerebrospinal fluid (CSF) flow artifacts (signal void, black arrows) due to its sensitivity to flow. In addition, CSF is clearly attenuated in the spinal cannel compared to bSSFP. CSF flow acts in a similar way as RF spoiling and thus results in T1-weighted dark CSF signal. Pure T1 contrast can be restored in RF-spoiled GRE (mid) accompanied by a loss in SNR. Balanced SSFP (right) provides high SNR and good fluid-tissue contrast but banding artifacts (yellow arrows) demonstrate its sensitivity to local field inhomogeneity. [Color figure can be viewed in the online issue, which is available at wileyonlinelibrary.com.]

Unbalanced GRE

After the readout process, additional “spoiler” gradients are used to effectively average over different dephasing states of the transverse magnetization. The net gradient area is nonzero but constant within each TR.

RF Spoiled GRE

The phase of the RF pulses is systematically incremented, leading to a steady state that is approximately T1-weighted according to the Ernst equation. As for unbalanced GRE, a spoiler gradient is necessary as well.

Balanced SSFP (bSSFP)

The gradients are arranged such that the net gradient area over one repetition time TR is zero along each axis. This results in an efficient “recycling” of all available magnetization for MR signal generation but with the undesirable result of increased sensitivity to field inhomogeneities.

A more generic name for the whole family of fast GRE sequences that is often used is “SSFP” (steady-state free precession) (31). This term refers to magnetization precessing freely (ie, without being exposed to RF) between consecutive RF excitation pulses. In this context, some confusion can result from nomenclature of these sequences, as there are different acronyms used in the literature and by different vendors for one and the same sequence. Moreover, some acronyms have changed their meaning over the years (eg, the acronym FISP (32)). The only advice that can be given is to read publications thoroughly and to not rely on the meaning of acronyms. An overview of fast GRE techniques and most common vendor-specific acronyms typically used for the different sequence types is also provided in the review article by Elster (33) and in Figure 10.

Unbalanced GRE

The simplest form of fast GRE imaging, unbalanced GRE, corresponds to the basic GRE pulse sequence with short TR (<T2) and “spoiler gradients” to provide sufficient dephasing of the transverse magnetization. In the example in Figure 6 (left), spoiling is simply accomplished by lengthening the readout gradient prior to application of the next RF excitation. Despite the spoiler effect, the transverse magnetization does not fully decay before the execution of the next RF pulse such that unbalance GRE provides high signal-to-noise ratio (SNR) due to coherent use of magnetization but offers complex T2/T1-contrast (as opposed to pure T1 contrast for basic GRE with TR ≫ T2) that may make image interpretation difficult. In addition, the complex interaction of nondecayed transverse magnetization over several TR intervals can result in flow-induced artifacts, as shown in Figure 6 (left, black arrows).

The mixed T2/T1-contrast and sensitivity to flow and motion artifacts are often problematic for clinical applications. As a result, unbalanced GRE does not play an important role in routine clinical applications. Nevertheless, the method represents an important basis for fast GRE imaging. To address the shortcoming of unbalanced GRE, two different fast GRE variants have been developed: RF spoiled GRE and balanced SSFP.

RF spoiled GRE

One of the key features of clinically used fast GRE has been the ability to provide well-defined T1 contrast while maintaining high imaging speed. To enhance T1 contrast while maintaining the short repetition times, TR, necessary for fast imaging, a solution is provided by the so-called RF spoiling technique (28, 34-38). As shown in Figure 6 (mid), the method is based on quadratically incrementing the phase of the RF pulse from TR to TR, eg, resulting in a phase ϕ(n) of the n-th RF-pulse of ϕ(n) = n(n-1/2, with a constant number ω. The phase of an RF pulse refers to the “tipping direction” when the magnetization is flipped from the longitudinal direction to the transverse plane and determines the orientation of the magnetization after the pulse in the transversal (x-y-) plane. It can be shown that for certain RF phase increments (eg, ω = 117° or 50°) the T2 contributions from transverse magnetization can effectively be eliminated. As a result, RF spoiled fast GRE sequences are predominantly T1-weighted. The signal intensity in RF spoiled GRE images effectively obeys the Ernst equation in the same way as for the basic GRE sequence. Note that RF spoiling does not destroy magnetization in the strict sense. Instead, RF spoiling manipulates magnetization such that the vector addition of all transverse magnetization components results in a zero net magnetization at the end of every sequence interval.

Figure 6 shows a side-by-side comparison of RF spoiled GRE and an unbalanced GRE acquisition. It is evident that RF spoiling can successfully be used to eliminate T2 contributions and enhance T1 contrast while maintaining short TR. It should be noted, however, that RF spoiled GRE carries an SNR penalty since the transverse magnetization does not contribute to the total MR signal.

RF spoiled GRE sequences are a common workhorse in the clinical routine and find a wide variety of applications for which the availability of T1-contrast in combination with short scan times is essential (eg, rapid imaging for large volumetric coverage). Important applications include tissue perfusion imaging (15, 16) or infarct imaging with delayed enhancement (13, 14). In cardiovascular MRI, frequently employed clinical applications include first-pass myocardial perfusion imaging and contrast-enhanced MR angiography (39), which rely on the T1 shortening of contrast agents in combination with fast imaging to capture the passage of an intravenous bolus injection (Figure 7). Historically, RF spoiling made fast T1-weighted imaging possible and opened the door to numerous contrast-enhanced clinical applications.

image

Contrast-enhanced (CE) MR angiography (MRA) and assessment of myocardial perfusion using fast RF-spoiled GRE imaging. a: Intravenous Injection of a Gd-based contrast agent results in T1-shortening in blood during the passage of the contrast agent bolus. b,c: The combination of appropriate synchronization of the measurement with the arrival of the arterial contrast permits the acquisition of the contrast agent bolus passage with high arterial signal and contrast. RF-spoiled 3D GRE is ideally suited for CE-MRA due to the possibility for rapid data acquisition with pure T1 contrast. CE-MRA can be applied in all vascular regions throughout the body as illustrated for cervical and cranial arteries (left), the thoracic aorta (mid), and hand (right). d: The first-pass arterial passage of contrast agent through myocardial tissue can monitored using ECG gated saturation recovery (SR) fast 2D RF spoiled GRE. In typical clinical applications, 3–5 short axis slices are acquired per RR-interval permitting the measurement of myocardial signal changes during contrast agent passage with a temporal update rate of one heartbeat. Prior to data acquisition for each slice, a 90° SR pulse is used to enhance contrast between enhancing and nonenhancing tissue. The two short axis images demonstrate an endocardial perfusion defect (arrows) during early (left) and late (right) arterial phases.

A remark on the term “gradient spoiling”: Cancellation of transversal magnetization components can also be established by spoiler gradients varying along the slice-selection direction from TR to TR (40). However, this “gradient spoiling” technique is less robust than RF spoiling; therefore, it is rarely used in recent clinical GRE imaging methods (21, 34). Gradient spoiling is conceptually different from the use of a constant spoiler gradient, which is used for both RF spoiled and unbalanced GRE. Note that in unbalanced GRE sequences, flow can also lead to T1-weighting, as it can act in a similar way as RF spoiling (eg, attenuation of flowing CFS in Figure 6 in the spinal canal, left).

Balanced SSFP (bSSFP)

A third fast GRE variant is known as bSSFP imaging (17) and has gained increased importance in a number of applications due to its superior SNR and in particular blood–tissue contrast (41, 42). Compared to other GRE variants, gradients are fully balanced (ie, gradient areas sum to zero over TR, see Figure 6, right), resulting in a more efficient refocusing of the steady state magnetization and thus increased signal.

However, bSSFP also exhibits a strong sensitivity to local field offsets such as variations of the external field or regional susceptibility differences (off-resonance sensitivity) (43). Signal and contrast for bSSFP imaging are a function of the local magnetic field. If the magnetic field is nonuniform, there can be regions in the image where the signal almost completely vanishes (banding artifacts, see Figs. 6, 8).

image

Time-resolved ECG-gated cardiac imaging with bSSFP and RF spoiled GRE. Improved image quality and enhanced blood–tissue contrast for bSSFP are clearly visible. Note the appearance of SSFP banding artifacts (arrows) visible as signal void in regions of increased field inhomogeneity. LV, left ventricle; RV, right ventricle.

In addition to the local field, the repetition time TR can substantially influence the spatial bSSFP signal homogeneity. As TR is increased, image quality degrades and banding artifacts in areas of high local field offsets become more and more prominent. A combination of good field homogeneity (the so-called “shim”) and short repetition times is therefore crucial for artifact-free bSSFP imaging. The short repetition times needed by bSSFP can only be achieved with sufficiently fast switching and strong gradients. This is the reason why bSSFP imaging did not become widely used until the late 1990s (17), when the necessary gradient performance became available. Balanced SSFP sequences are mostly used for, but not limited to, body and cardiovascular MRI (17, 42).

In addition to local field inhomogeneities, bSSFP is sensitive to any disruption of the steady state, ie, by interruption of the regularly spaced train of RF pulses. For applications that rely on repetitive magnetization preparation or added functionality, such as fat saturation, tagging preparation, or navigator gating, the steady state must be interrupted to include an appropriate preparation sequence. In order to avoid severe disruption of the steady state magnetization, the RF pulses and gradients necessary for the preparation are typically embedded into a steady state storage scheme (“α/2 technique”) as presented by Scheffler et al (44).

Both RF spoiled GRE and balance SSFP are used for dynamic cardiac imaging. If data acquisition is gated or synchronized with the cardiac RR interval, time-resolved (CINE) anatomical images can be collected to depict the dynamics of left and right ventricular function during the cardiac cycle (45). For short TR, the superior signal amplitude and improved blood-tissue contrast of bSSFP is obvious from Figure 8. However, sensitivity of bSSFP to field inhomogeneities may result in signal variations and artifacts. The effects become more prominent at higher fields (3 T and higher) such that cardiac imaging with balanced SSFP has become the method of choice at 1.5 T, while RF spoiled GRE, which is less sensitive to local field variations, is often used at 3 T. Figure 8 shows results from an electrocardiogram (ECG)-gated dynamic study of cardiac motion in the four-chamber view. Note the high SNR and also blood–tissue contrast of bSSFP (bottom row) as compared to RF spoiled gradient echo imaging (top row).

Comparison of Fast GRE Techniques

Figure 9 provides an overview of the signal and contrast characteristics for all three fast GRE techniques and basic GRE as a function of the flip angle used for RF excitation. Simulations based on Bloch equations have been employed to calculate signal and contrast of brain white matter (WM) and gray matter (GM). Relaxation times at 1.5 T (20) are: WM: T1 = 600 msec, T2 = 80 msec; and GM: T1 = 950 msec, T2 = 100 msec. Three different repetition times were simulated: TR = 6 msec (≪ T2); TR = 60 msec (≊T2); TR = 180 msec (>T2, <T1). Contrast is simply the difference in signal amplitudes between WM and GM.

image

Simulation of signal (left column) and gray matter/white matter contrast (right column) for various GRE sequences.

RF spoiled GRE shows signal behavior according to the Ernst equation, with maximum GRE signal at the Ernst angle. The maximum shifts to lower flip angles with shorter TR. Note that at the Ernst angle all three sequence variants exhibit the same signal intensity. For longer TR, the signal curves become more and more similar, with maximum signal intensity shifting closer to 90°. Contrast increases with longer TR, and is highest for RF spoiled GRE for intermediate and long TR. Simulations confirm the finding from Figure 5, contrast of RF spoiled GRE is maximized for flip angles higher than the Ernst angle (26, 27).

The basic properties and commonly used acronyms of the three most widely used fast GRE techniques are summarized in Figure 10. Unbalanced GRE demonstrates high signal but rather low and complex T2/T1-contrast and is sensitive to flow and motion artifacts. T1 contrast is enhanced by the application of RF spoiling but overall signal intensity is considerably reduced. Balanced SSFP imaging can offer the highest signal but suffers from strong sensitivity to local field changes and also provides mixed T2/T1 contrast (43).

image

Fast gradient echo imaging variants, fundamental signal and contrast properties and corresponding vendor-specific acronyms.

T*2 and Off-Resonance Effects

One of the fundamental differences between GRE and spin echo imaging is related to the fact that phase shifts from local magnetic field inhomogeneities are not refocused in GRE imaging and lead to signal decay governed by T*2.

In general, GRE images are thus more sensitive to artifacts due to changes in magnetic susceptibility (46). For long echo times, signal loss can be severe, as shown by the GRE images in Figure 11, where it is demonstrated that a longer TE leads to an overall decline in signal intensity and almost complete signal dropout in the frontal region with increased field inhomogeneity resulting from high susceptibility differences. For comparison, spin echo imaging is independent of these effects.

image

Turn:x-wiley:10531807:media:JMRI23638:tex2gif-stack-3-effects in GRE imaging illustrated for basic multislice 2D GRE (TR ≫ T2) for two images at different axial location in the head. The GRE images are compared to spin echo (SE) images at identical anatomical locations using the same echo and repetition times. For short TE = 4.1 msec, both GRE and SE images are predominantly proton density (PD)-weighted, show similar contrast, and no or only minor artifacts. Increasing echo time to TE = 30 msec, GRE images show overall signal reduction and marked signal loss in regions which have high susceptibility variations (air–tissue interfaces, white arrows). In contrast, spin echo imaging is insensitive to these effects. Note that long echo times result in T2 contrast for SE images while GRE images are Turn:x-wiley:10531807:media:JMRI23638:tex2gif-stack-4-weighted. All images were acquired at 1.5T.

However, T*2-sensitivity of GRE imaging can also be beneficial and used to generate new diagnostic information, as shown in Figure 12. For example, T*2-contrast is used in functional MRI (fMRI) where changes in blood oxygenation that lead to alterations in T*2 are used to analyze brain activation (Figure 12b) (47, 48). Another widely used application involves the detection of bleeding in stroke patients. Microbleeds in such patients lead to local field alterations and therefore T*2-changes that can be visualized as signal dropouts using GRE techniques (49) with long echo times (Figure 12b). Since such bleeds are very small and hardly detectable with any other imaging modality, T*2-GRE imaging is often part of the clinical routine workup of stroke patients. Furthermore, T*2 differences are also exploited in susceptibility-weighted imaging (SWI) (50). Additional applications include the analysis of the T*2-dynamics in contrast agent studies (51) that can be used to generate brain perfusion maps (52, 53) such as regional cerebral blood volume and flow.

image

Examples of the application of Turn:x-wiley:10531807:media:JMRI23638:tex2gif-stack-5-weighted gradient echo imaging. a: Cranial 2D GRE with long echo times reveal hemosiderin deposits as a result of microbleeds (white arrows) after stroke. b: Functional MRI used GRE with long echo times to detect changes in regional blood Turn:x-wiley:10531807:media:JMRI23638:tex2gif-stack-6 based on to changes in local blood oxygenations levels related to neuronal activity (BOLD effect). [Color figure can be viewed in the online issue, which is available at wileyonlinelibrary.com.]

A second important effect acting on the observed signal intensity of gradient echo sequences is the chemical shift. Due to the frequency difference of water and fat (220 Hz at 1.5 T, 440 Hz at 3 T), their magnetization vectors exhibit a phase difference, which depends on the echo time TE. Of special importance are the in-phase (fat and water magnetization are aligned and add) and the out-of-phase (fat and water magnetization are opposed and cancel) conditions. For a 1.5 T magnet, images can be produced that provide contrast generated by opposed (TE = 2.2 msec, 6.6 msec,…) and aligned fat and water spins (TE = 4.4 msec, 8.8 msec,…).

Figure 13a shows an example of in-phase/out-of-phase images, which can be used to analyze fat and water content in different tissues (54). Since fat and water signal add in the in-phase image, signal intensities are typically higher than in the out-of-phase images. Note the occurrence of dark rims at the interface between fatty and aqueous tissues when TE is chosen to correspond to the out-of-phase condition. The signal void represents partial volume effects in pixels with roughly equal fat and water content which cancel and lead to signal void. Furthermore, the information in both images can be combined and processed for effective fat–water separation (55-57), eg, for imaging of joints for cartilage analysis (58), diagnosis of fatty liver disease (59), or lesion characterization as shown in Figure 13b.

image

a: Principe of fat water imaging based on in-phase and out-of-phase images for GRE imaging with different echo times TE. Fat and water images calculated for the in-phase/out-of-phase data. b: Cardiac fat-water imaging in four-chamber orientation clearly showed intraarterial lipomatous signal as visible in the fat image and absent in the water image (white arrows). Fat-water imaging permitted an improved characterization of this lesion compared to standard GRE imaging (left).

This manifestation of the chemical shift must not be confused with the so-called “chemical shift misregistration” artifact (22), which occurs in GRE as well as in SE sequences. In this context, chemical shift can cause a spatial misregistration of fat versus water tissue depending on the bandwidth of the pulse sequence.

Flow Effects

Since signal intensity in GRE imaging is determined by repeated RF excitation, any static tissue will typically not fully recover prior to the next excitation (steady state formation), resulting in saturated and consequently reduced signal intensities. Then signal behaves according to the steady state simulations for which typical examples have been shown in Figure 9. In contrast, blood that flows into an imaging slice has not experienced a sufficient number of previous RF excitations to build up a steady state. This will lead to signal enhancement, ie, in-flow of unsaturated magnetization results in bright signal of blood as typically observed in fast GRE imaging.

Such inflow enhancement can be beneficial and used to acquire additional diagnostic information as, for example, shown in Figure 14 for time-of-flight (TOF) techniques (60, 61). The in-flow enhancement is used to separate arterial or venous signal from static tissue or background signal to generate angiograms. Here the optimal flip angle for high blood/tissue contrast is usually higher than for static tissues such as the example of gray and white matter shown in Figure 5. Additionally, background (static tissue) suppression techniques such as magnetization transfer pulses can be employed (62).

image

In-flow enhancement in GRE imaging results in bright signal of blood entering the imaging slice. In combination with saturation of static tissue this effect can be exploited to provide high blood-tissue contrast as in TOF MR angiography. [Color figure can be viewed in the online issue, which is available at wileyonlinelibrary.com.]

In-flow can also lead to image artifacts such as misregistration and ghosting (22) that arises if laminar or pulsatile flow leads to amplitude or phase inconsistencies. A solution to this problems is offered by flow compensation techniques (63), which can eliminate artifacts but lead to an increase in echo and repetition times TE and TR.

Magnetization Preparation

The intrinsic contrast of GRE imaging techniques may not be sufficient, for example, when using short TRs. In this case, as can be seen from Figure 9, signal and contrast are rather poor for unbalanced and RF spoiled GRE. A potential solution is offered by magnetization preparation methods, which are used to imprint T1 or T2 contrast onto the longitudinal magnetization Mz and enhance the desired contrast and signal amplitude. A frequently used technique is shown in Figure 15, which illustrates enhanced T1 weighting by inversion recovery imaging (64). Before the actual start of GRE imaging, a 180° inversion pulse is applied preceding data acquisition by the time-interval or inversion time TI. Depending on T1 of the tissue and the TI time, certain T1 species can be enhanced, suppressed, or even nulled. For cranial imaging, magnetization prepared rapid gradient echo (MP-RAGE) is often used clinically for anatomical imaging of the entire head with T1-contrast. Another clinically important example of an inversion recovery application is the “delayed hyper-enhancement” technique (65), which is used to identify infarcted and nonviable tissue in patients with coronary artery disease. Here, the inversion time TI is selected such that the myocardial signal is nulled and only contrast agent that accumulates in infarcted tissue—and has much shorter T1—appears bright in the GRE images (Figure 15).

image

Left: Magnetization preparation by inversion recovery (IR) for enhanced T1 contrast as in delayed enhancement cardiac MRI. To improve contrast between tissue with and without contrast agent uptake, the inversion time (TI) can be selected to null nonenhancing myocardial tissue (long T1). Contrast agent that accumulates in infarcted tissue results in shorter T1 and high GRE signal. Right: Delayed enhancement images of the left ventricle (LV) in a patient with myocardial infarction. Bright areas (arrows) mark regions with contrast agent accumulation in irreversibly damaged myocardial tissue.

DISCUSSION

In this review the basic principles of GRE imaging and the most commonly used techniques for fast GRE data acquisition have been presented. A brief summary of GRE properties is provided in Figure 16. Noticeably, the signal of a gradient echo sequence is sensitive to various T*2-effects, which—contrary to spin echo acquisition—are not refocused at the time of signal detection.

image

Summary of the properties of gradient echo imaging.

Signal and contrast for basic GRE imaging (TR ≫ T2) depend on the formation of a steady state that is governed by the Ernst equation. T2-contribution can be ignored, resulting in relatively simple contrast behavior depending on TR, T1, and the flip angle. Due to the long TRs (on the order of several 100 msec), basic GRE does not offer many advantages over spin echo imaging, which often provide improved image quality for similar total scan times. Basic GRE, therefore, does not play an important role in clinical applications. In contrast, fast GRE techniques provide substantially higher imaging speed and are widely used in numerous applications in the clinical routine.

Fast GRE imaging techniques (short TE and TR ≪ T1 and < T2) demonstrate complex contrast behavior depending on TR, T1, T2, flip angle, and in case of bSSFP, even the local magnetic field (off-resonance effects). Unbalanced and balanced GRE techniques provide complex T2/T1 contrast and high SNR. Balanced SSFP offers even higher signal and especially excellent blood tissue contrast but suffers from sensitivity to local field homogeneity. T1 contrast can be restored by the application of RF spoiling, which offers a robust and fast imaging technique for a variety of applications.

All the discussed GRE sequences rely on the principle of SSFP. It is important to note that the family of SSFP sequences contains even more members, such as PSIF (also called CE-FAST) (66, 67) or echo-shifted GRE sequences (68); all of them are distinguished by the arrangement of gradient pulses. However, in clinical practice these SSFP variants play only a very minor role, and are thus not covered by this review article. The reader is referred to further literature for more information (29, 35).

An important extension of GRE techniques are ultrafast GRE methods, which combine gradient echo formation with multiecho data acquisition or time-efficient sampling strategies such as spiral (69-71) or echo planar imaging (72, 73). A more detailed description and discussion of the properties and methodological considerations of these techniques is beyond the scope of this article. More details can be found in a recent review article by Tsao (74).

In summary, GRE sequences play an important role in MRI due to several reasons: First, the build-up of a steady state enables fast imaging and 3D acquisitions. Second, the broad variety of contrast behavior and sensitivity to other influences such as flow makes it possible in a large number of cases to find a suitable gradient echo sequence as the matter of choice. Third, due to their robust and reliable image quality, GRE sequences can be used as acquisition modules in combination with various magnetization preparation modules.